FDA Comment
Considerations for the Regulation of Generative AI-Enabled Medical Devices
Circuit Breaker Labs | FDA Public Comment • October 19, 2026
Assessing Risk
Question 1: Risk Framework
Does the proposed two-axis risk framework adequately capture the risks of generative AI medical devices, or should additional factors be considered?
The proposed two-axis framework provides a useful foundation, particularly because it recognizes risk as a continuum. However, device activity and the consequences of an incorrect output do not fully characterize the risks of generative AI-enabled medical devices. We recommend retaining the two axes while introducing additional axes.
First, risk depends on the characteristics of the intended user. A patient, a practicing clinician, a specialist, and a clinician in training may interact with the same system differently and have different abilities to recognize an incorrect response. Even within these populations, age, health literacy, clinical complexity, and patterns of use can materially affect risk. For instance, adolescent users may use language that differs substantially from the system's expected inputs. Risk assessment should consider foreseeable use behaviors rather than assuming all users interact with the device as intended.
Second, informational functions warrant greater differentiation. A resource navigator that directs users to established clinical information differs meaningfully from a system that synthesizes medication information in the context of an individual's treatment history. Even without an explicit recommendation, personalized information may influence clinical decisions. The extent to which a system incorporates patient-specific information should therefore modify its risk classification.
Third, the presence of human supervision should not automatically be treated as equivalent to effective oversight. A clinician who reviews every output before it reaches a patient provides a different safeguard from one who retrospectively reviews a small sample of interactions. Risk assessment should consider the timing, frequency, coverage, clinical competence, and intervention authority associated with human oversight. Retrospective sampling may detect systematic problems but cannot prevent harm from an individual output already delivered.
Fourth, traceability and robustness should be considered together. A system may correctly handle a clinically significant concept when expressed using formal terminology but respond differently when the same information is expressed colloquially. For example, requests involving “acetaminophen” and “Tylenol” should not produce materially different safety-critical behavior merely because of the terminology used, but we have observed materially different behavior in our evaluations at Circuit Breaker Labs. Traceability to appropriate clinical sources is valuable, but it does not itself establish that the system will apply those sources consistently.
Rather than adding numerous independent axes, CDRH could establish a small set of risk modifiers and associated testing expectations. This would preserve the flexibility of the proposed framework while giving manufacturers more predictable, application-specific requirements.
Question 2: Informational Functions
How should the FDA assess the risks of AI-generated information based on its wording, personalization, context, and influence on user decisions?
The degree to which an informational function directs behavior cannot be assessed solely through explicit instructions or recommendations. Risk also depends on how the information is communicated, the user's circumstances, and the degree of authority or interpersonal trust the system may implicitly convey.
A particularly important consideration is anthropomorphism, the attribution of human characteristics or understanding to an AI system. A system that speaks in a highly personal, emotionally affirming, or companion-like manner may encourage users to place greater confidence in its outputs, even when those outputs are nominally non-directive. Generative systems may also mirror or reinforce a user's beliefs and emotions, a behavior that can be inappropriate in certain clinical contexts.
We recommend a principle of minimum necessary anthropomorphism: manufacturers should justify personifying design features according to their intended clinical benefit and assess whether those features introduce undue trust, emotional dependence, or inappropriate behavioral influence. This need not prohibit first-person language or empathetic communication. Indeed, some therapeutic applications may benefit from carefully designed relational interactions. The important distinction is between communication that supports appropriate engagement and communication that encourages users to overestimate the system's understanding, competence, or authority.
CDRH could provide greater predictability through a behavioral assessment framework considering the use of patient-specific information, implied authority, emotional influence, and persistence of advice across repeated interactions.
We also recommend that manufacturers evaluate these properties across relevant patient populations, clinical conditions, and conversational contexts. Attachment-related risks may emerge gradually, particularly during extended interactions, and may not be apparent from isolated responses.
Rather than requiring manufacturers to eliminate all personalization, CDRH should encourage empirical evaluation of how personalization affects comprehension, trust calibration, and downstream decision-making.
Question 3: Patient-Facing AI
How should the FDA account for the unique risks of delivering AI-generated clinical information directly to patients rather than clinicians?
Patient-facing generative AI applications offer substantial potential benefits, including more timely access to health information, improved patient understanding, and support when a clinician is not immediately available. These benefits should not be discounted merely because patients lack formal clinical training.
However, patient-facing systems may require stronger safeguards because the recipient may not be able to identify an incorrect output or recognize when an apparently routine question involves clinically significant risk.
We recommend an approach in which multiple complementary safeguards address different failure modes. These should include clinically-grounded constraints, appropriate uncertainty communication, reliable escalation pathways, and testing against foreseeable real-world inputs.
An important consideration is the reliability of the safeguards themselves. Some systems use generative language models to evaluate or moderate the outputs of other generative models (‘LLM-as-a-judge’). Such approaches can introduce additional model-dependent error or variability. For safety-critical judgments, manufacturers should characterize the evaluator's sensitivity, specificity, reproducibility, variance, and susceptibility to adversarial or out-of-distribution inputs.
Patient-facing systems should also be evaluated using realistic input variation, including misspellings, colloquialisms, incomplete medical histories, contradictory information, differences in health literacy, and population-specific language. The goal should be reliable performance across the intended population.
Clinical claims should be supported by appropriate, current evidence, with uncertainty clearly communicated where evidence is limited. In clinically ambiguous or high-consequence circumstances, systems should be able to defer or escalate appropriately without over-refusing routine informational requests.
Importantly, these safeguards should be assessed through observable behavior, not solely through documentation that a safety policy or guardrail has been implemented.
Question 4: Clinical Specialization
How should differences between generalist and specialist physicians affect the risk assessment and safeguards of clinician-facing AI?
The distinction between generalist and specialist clinicians is important, but risk assessment should reflect the actual clinical workflow. We recommend involving both generalist and specialist clinicians in the development and evaluation of relevant GenAI-enabled functions. Generalists can help assess how a system is likely to be used in routine practice, while specialists can identify subtle errors, contraindications, and clinically significant failure modes that might otherwise go undetected.
This is particularly important for systems that provide specialized information to users without corresponding specialist training. A physician may appropriately rely on an AI system for support outside their primary field but may be less equipped to recognize when its recommendations are subtly incorrect.
Behavioral health provides a useful example. An AI system that repeatedly reassures or validates a user's concerns may appear an empathetic sounding board. However, in the context of obsessive-compulsive disorder, repetitive reassurance can reinforce maladaptive cycles. Similarly, uncritical affirmation of delusional beliefs may be clinically inappropriate. Evaluations developed without relevant specialist input may fail to distinguish appropriate emotional support from harmful reinforcement.
We recommend that clinical specialists contribute not only to reviewing outputs, but also to defining failure categories, evaluation criteria, and clinically appropriate safeguards. Generalist clinicians should participate in realistic workflow testing to determine whether the system communicates these distinctions effectively to its actual users.
For applications whose intended use population could reasonably span multiple clinical domains, the evaluation should also include relevant adjacent specialties and common comorbid conditions. A device intended for one specialty may create risks outside that specialty through medication interactions, overlapping symptoms, or treatment decisions.
Question 5: Multi-Turn Conversations
How should the FDA evaluate risks that emerge or change over the course of extended AI conversations?
Multi-turn conversational devices introduce risks that cannot be adequately characterized through isolated prompts. Safety-critical behavior may change over the course of a conversation as information accumulates, earlier details become less salient, and the user's presentation evolves.
One particularly important concern is context pollution. Clinically consequential information introduced early, e.g., during patient history, may no longer influence later outputs appropriately because of context-window limitations, summarization errors, retrieval failures, or the model's inconsistent use of available information.
For example, a patient may disclose a serious medication allergy early in an interaction, then return to medication-related questions much later. The continued absence of an allergic reaction does not make that contraindication irrelevant. Similarly, a system instructed to provide non-directive information may gradually transition toward individualized treatment recommendations as conversation context accumulates.
We recommend that CDRH encourage evaluation of complete conversational trajectories, including scenarios in which risks emerge gradually, relevant information is disclosed indirectly, users resist escalation, and important details are separated by substantial conversational distance.
Testing should examine whether safety-critical constraints persist despite changes in topic, user framing, and conversation length. Manufacturers should characterize the reliability of their memory mechanisms, context-management strategies, and safety controls under realistic and adversarial conditions.
Where appropriate, safety-critical information should be preserved in structured, persistent representations rather than relying exclusively on the generative model to recall and correctly prioritize it. Such representations should themselves be tested for accuracy, appropriate updates, and protection against conflicting or misleading information.
We further recommend distinguishing turn-level safety from trajectory-level safety. A system may produce individually plausible responses while its cumulative behavior becomes unsafe. The unit of evaluation must therefore include the conversation and its clinical consequences, not merely each response in isolation.
Question 6: Care Escalation
How should manufacturers measure and balance the risks of escalating care too frequently versus failing to escalate when necessary?
Under-escalation and over-escalation should be characterized separately because their consequences differ across clinical contexts and cannot always be reduced to a common error measure.
Repeated unnecessary escalation may also discourage patients from engaging with the system or seeking appropriate care in the future.
We recommend defining escalation thresholds according to the severity of potential harm, time sensitivity, uncertainty, availability of clinical supervision, and reversibility of an incorrect decision.
For instance, a clinician-supervised informational function may appropriately provide differentiated risk information without reflexively escalating every ambiguous presentation. Conversely, an autonomous function operating without timely human review may require a lower threshold for escalation in high-consequence situations.
Clinical prevalence and patient heterogeneity should also inform evaluation. In rare or highly variable conditions, the absence of a familiar symptom pattern may provide limited reassurance. However, rarity alone should not automatically mandate escalation. The relevant consideration is whether uncertainty and potential harm justify additional review.
CDRH should encourage manufacturers to report under-escalation and over-escalation as distinct outcomes, with clinically justified thresholds for each. Evaluations should include true emergencies, borderline presentations, and clearly non-urgent cases.
Particular attention should be given to escalation timing. A system that ultimately recommends emergency care may still fail if it first provides several rounds of inappropriate reassurance or delays the recommendation beyond a clinically meaningful point.
Competency-Based Premarket Evaluation
Question 7: Competency-Based Evaluation
Is a framework combining device benchmarking with clinical confirmation appropriate for evaluating generative AI medical devices?
We support the proposed competency-based approach, particularly the distinction between scalable non-clinical benchmarking and clinical confirmation under representative conditions.
That said, representative, high-volume testing of open-ended generative systems is not impractical. As a point of evidence: Circuit Breaker Labs has designed an evaluation system for monitoring of clinical AI that can run hundreds of thousands of dynamic, multi-turn tests each day, building meaningful clinical evidence at rapid speed.
Importantly, competency should not be understood as a static property established at the time of initial evaluation. Generative AI-enabled devices may change because of modifications to underlying models, prompts, guardrails, retrieval systems, user interfaces, or other components. Even changes not explicitly intended to affect clinical or safety performance may alter behavior in a clinically significant way.
We therefore recommend treating competency-based evaluation as the foundation of a continuous assurance process. Premarket benchmarking would establish a characterized baseline, clinical confirmation would provide evidence of performance in representative use, and postmarket reassessment would determine whether that competency is maintained as the system evolves.
The core competencies and acceptance criteria remain stable to support meaningful longitudinal comparisons. At the same time, individual test cases should be refreshed and expanded to account for emerging failure modes, changing clinical practices, and previously unidentified populations or conditions.
This would preserve standardization without allowing initial benchmark performance to become a substitute for continued evidence of safety and effectiveness.
Question 8: Risk-Based Evidence
How should a device's risk classification determine the amount and type of evidence required before FDA approval?
The two-axis framework could provide the foundation for determining the rigor and scope of premarket evidence, but the additional risk modifiers discussed in Question 1 should also inform what manufacturers are expected to demonstrate.
In particular, CDRH should consider the intended user, degree of personalization, clinical supervision model, patient population, potential for extended interactions, and clinical dependencies extending beyond the device's primary specialty.
A device providing general cardiovascular education, for example, should not necessarily face the same evidentiary expectations as one generating individualized cardiovascular treatment recommendations. Likewise, a cardiology-focused device may need to demonstrate appropriate handling of pulmonary disease, renal impairment, and other relevant comorbidities when they influence its intended function.
We recommend that manufacturers establish a risk-to-evidence mapping before testing. This would identify which clinically significant behaviors must be evaluated, which populations and deployment conditions must be represented, and what acceptance thresholds are justified for each category.
This mapping would also provide a reproducible starting point for postmarket evaluation. When device components change, manufacturers could determine which risk categories may have been affected and conduct targeted reassessment, while retaining broader regression testing for unexpected effects.
A standardized approach to mapping risk categories to evidence requirements could improve regulatory predictability without imposing identical testing obligations on devices with materially different risk profiles.
Question 9: Benchmarking Standards
Does the proposed benchmarking framework adequately assess clinical knowledge, reasoning, safety, communication, and generalizability, or are additional standards needed?
The proposed benchmarking elements provide a strong foundation. We particularly support the inclusion of safety-critical escalation, scope adherence, uncertainty communication, robustness, subgroup performance, and agentic capabilities.
However, we recommend some refinements.
First, CDRH should distinguish reusable evaluation criteria from reusable test instances. Standardized clinical competencies, scoring definitions, and acceptance criteria are essential for comparability. Repeated reliance on a fixed, publicly available set of prompts or data is less reliable, particularly when models or applications are often optimized against those exact examples. The evaluation framework should therefore support standardized evaluation metrics applied to both reference cases and independently generated, sequestered, or refreshed scenarios.
Second, clinically significant context retention should be an explicit competency. Devices should demonstrate that they maintain and appropriately apply relevant patient information across extended interactions, including contraindications, allergies, comorbidities, and changes in clinical status. This competency spans clinical reasoning, safety, robustness, and agentic behavior, but its importance warrants explicit consideration.
Third, anthropomorphism, inappropriate agreeableness, and automation bias should receive greater attention. A system may provide factually accurate information while encouraging excessive reliance or reinforcing harmful beliefs. These risks can emerge through communication style, emotional affirmation, and repeated interactions, particularly in behavioral health applications.
Fourth, clinical provenance should be assessed. Where a device presents information as clinically established, users should be able to distinguish established guidance from emerging evidence, inference, or uncertain claims. Unsupported assertions of authority should be considered separately from ordinary factual errors.
Finally, the proposed elements overlap in clinically meaningful ways. For example, an agent that reorders a medication without adequately considering a relevant contraindication may demonstrate failures of clinical proficiency, context handling, and agentic execution simultaneously.
We do not recommend treating these categories as mutually exclusive. Instead, CDRH could retain the organizing structure while allowing individual failures to map to multiple competencies, with careful safeguards against double-counting in aggregate scores.
Question 10: Benchmark Validity
How can manufacturers demonstrate that benchmark performance reliably predicts real-world safety without overfitting or introducing conflicts of interest?
CDRH identifies a central challenge in generative AI evaluation: performance on a known benchmark does not necessarily predict performance under novel, real-world conditions.
We recommend distinguishing the clinical construct being measured from the particular examples used to measure it. Clinical risk categories, evaluation criteria, and acceptance thresholds should be clearly specified and sufficiently stable to support reproducibility. However, the scenarios used to assess those should be capable of evolving as new failure modes and patterns of use emerge.
A robust approach could combine common reference benchmarks, independent sequestered cases, and dynamically generated scenarios tailored to a device's intended use. This would allow manufacturers to assess established competencies while reducing overreliance on a finite set of recognizable inputs.
The construct validity of such benchmarks should be supported through qualified clinical review, explicit links to clinically meaningful failure mechanisms, testing across realistic input variation, and comparison with independent clinical adjudication where appropriate.
We also recommend evaluating measurement reliability separately from model performance. When a scoring method produces inconsistent judgments across equivalent inputs or repeated assessments, an apparent change in device performance may reflect measurement variation rather than an actual change in safety. Manufacturers should characterize the repeatability, agreement with qualified adjudicators, and limitations of their evaluation methods.
Independent evaluation organizations can help identify blind spots arising from sponsor-developed test suites, particularly when they collaborate with clinical specialists. However, independence should not preclude device-specific customization. A useful structure could include a common set of clinically defined risks supplemented by independently developed scenarios addressing the manufacturer's intended use, population, and known vulnerabilities.
Manufacturers should document what risks their systems may encounter, why those risks matter, and independent evaluators can assess handling of these risks and provide detail on how the scenarios represent intended use, and what important uncertainties remain.
Finally, benchmark maintenance should be treated as a continuing scientific activity. As underlying models and clinical applications evolve, evaluation methods must be updated without sacrificing reproducibility or longitudinal comparability.
Question 11: Clinical Confirmation
How should manufacturers select appropriate clinical validation methods based on device risk, intended use, and real-world patient interactions?
We support a risk-proportionate approach to clinical confirmation that allows manufacturers to select methods appropriate to their device's intended use, degree of autonomy, and potential consequences of failure. However, we recommend that the selection process begin with a structured characterization of clinical risks rather than a determination of which study design is most convenient.
Manufacturers could first describe their device's risk profile using the proposed two-axis framework, supplemented by the additional considerations discussed in our earlier responses. This assessment should identify both the situations the device is intended to address and foreseeable circumstances in which its outputs could cause harm.
For example, a cardiology-focused system may not explicitly address pulmonary disease, yet many patients presenting with cardiovascular concerns also have respiratory conditions or overlapping symptoms. A risk assessment that excludes these clinical intersections could overlook consequential failure modes. CDRH could request additional evidence in areas where the intended function creates foreseeable dependencies on adjacent clinical specialties.
We recommend that manufacturers document these anticipated risks and provide corresponding performance evidence, including findings from qualified independent evaluators where appropriate. Independent evaluators can help identify blind spots that may be overlooked when manufacturers develop their own evaluation criteria, while clinical specialists and generalist physicians can provide complementary perspectives on clinical complexity and actual patterns of use.
The anticipated distribution of real-world inputs should inform both test coverage and the amount of testing performed. A device expected to support large volumes of patient interactions should be evaluated across a sufficiently broad range of scenarios to characterize uncommon but clinically significant failures.
High-throughput automated testing makes it feasible to examine many more scenarios than traditional manual evaluation alone.
For perspective, our work at Circuit Breaker Labs has focused on scalable, clinically-informed testing methods capable of evaluating large numbers of interactions efficiently. Such methods can supplement clinical confirmation by systematically examining variation in patient characteristics, language, comorbidities, and conversational trajectories.
It is also important to distinguish reproducibility of the evaluation method from variability in the device being evaluated. Reproducible scoring methods can make longitudinal comparisons more reliable, but their clinical validity must still be established through appropriate expert agreement and other evidence.
Finally, we recommend a staged approach in which scalable benchmarking identifies areas of uncertainty that then inform more resource-intensive confirmation methods. For instance, failures involving complex comorbidities discovered through synthetic testing could guide targeted retrospective analysis, standardized patient interactions, or shadow deployment.
This would allow manufacturers to use clinical evidence more efficiently while directing greater scrutiny toward the conditions most likely to produce significant harm.
Question 12: Statistical Reliability
How should manufacturers establish statistically meaningful performance estimates when combining synthetic testing, real-world data, and clinical evidence?
Statistically meaningful performance measurement is particularly challenging for generative AI-enabled devices because outputs may vary across repeated interactions, clinically significant failures may be uncommon, and the real-world distribution of patient inputs may be incompletely characterized.
We recommend that CDRH encourage large-scale, risk-stratified evaluation supported by clinically-informed synthetic data, while clearly distinguishing performance under synthetic testing conditions from estimated real-world performance.
Real-world patient data is essential for understanding how individuals actually interact with medical AI, including the conditions they present with, their language patterns, and the frequency of different clinical scenarios. However, particularly for emerging conversational and patient-facing applications, available data may not adequately represent the full range of potential interactions.
Privacy-preserving analysis of real-world data can help characterize these patterns and inform the construction of synthetic test scenarios. Clinical experts can then assess whether the generated cases are medically plausible, reflect meaningful patterns of presentation, and cover relevant populations and conditions.
Importantly, synthetic data should not merely reproduce the observed distribution of existing patient interactions. Real-world datasets may systematically underrepresent rare conditions, underserved populations, atypical symptoms, or clinically important scenarios that have not yet been encountered. Synthetic testing creates an opportunity to intentionally examine such situations before they result in real-world harm.
Advances in automated evaluation have made testing at the scale of hundreds of thousands or millions of interactions technically feasible, as evidenced by the in-market tool developed by Circuit Breaker Labs. We believe this capability should be leveraged to characterize performance across a broad range of clinically meaningful scenarios, including variations in patient language, age, comorbidities, and conversational histories.
We recommend separating three components of performance evidence:
We also recommend reporting uncertainty intervals, clinically relevant subgroup performance, and failure severity rather than relying exclusively on a single aggregate score.
The objective should be to make statistically rigorous evaluation more accessible and scalable.
Question 13: Synthetic Data
Where is synthetic data most useful for evaluating medical AI, and how can manufacturers prevent it from reproducing existing model biases and blind spots?
We believe synthetic data can make substantial contributions across clinical domains, device functions, and patient populations, particularly where real-world data is insufficient to characterize the full range of foreseeable inputs.
Conversational and behavioral health applications illustrate both the need for synthetic data and the importance of developing it carefully. Unlike many structured clinical datasets, patient-facing conversations can include indirect disclosures, evolving symptoms, incomplete histories, informal language, and substantial variation in how individuals describe similar experiences. Two patients presenting with the same underlying clinical concern may use entirely different language, and relatively small wording differences may influence whether an AI system recognizes a safety-critical condition or responds appropriately.
For example, a patient describing significant psychological distress may not use formal diagnostic terminology or explicitly identify the condition they are experiencing. A system evaluated primarily on textbook descriptions may perform well on a benchmark while failing to recognize clinically meaningful variations encountered in practice.
Synthetic testing offers a way to systematically explore these differences. It can generate variations in wording, age-related language, clinical context, comorbidities, symptom progression, and conversation length while preserving a clinically specified underlying scenario.
However, the clinical risks being tested should not be done solely by the same class of generative models being evaluated (‘LLM-as-a-judge approaches’). Otherwise judging may reproduce overlapping assumptions or blind spots.
Clinical experts should establish the relevant risk categories and behavioral expectations. Synthetic scenarios can then be generated through complementary techniques, including structured simulation and steering. Finally, outputs should demonstrate reliability and clinical validity.
At Circuit Breaker Labs, our approach emphasizes clinically-informed risk criteria and reproducible evaluation methods rather than solely on one generative model to independently produce scenarios and another to judge the results. We believe this separation is an important design principle for developing trustworthy evaluation infrastructure.
The involvement of clinicians experienced with underrepresented populations is especially important. Available patient datasets may not fully reflect how individuals across age groups, dialects, health literacy levels, cultural contexts, and co-occurring conditions describe their concerns. Specialists and generalists can identify clinically meaningful variations that would otherwise be absent from test-generation processes.
We also recommend that synthetic datasets be continuously expanded and validated as new evidence emerges from clinical practice, postmarket monitoring, patient feedback, and independent evaluations.
The appropriate goal is therefore not to replace real-world evidence, but to extend the range of conditions under which safety can be examined before patients encounter those conditions themselves.
Question 14: Performance Comparators
What clinical standards, physician expertise, or human-AI performance benchmarks should determine whether an AI device performs acceptably?
We recommend that performance comparators and acceptance criteria be determined according to a device's intended use, level of autonomy, patient population, and clinical consequences of failure.
The proposed axis framework could provide a starting point for establishing baseline expectations. Manufacturers should define the functions within the device's intended scope, the behaviors that would constitute unacceptable deviations from that scope, and the clinical competencies required for safe operation.
Minimum requirements could include recognition of clinically significant contraindications, appropriate escalation in emergencies, adherence to the device's intended scope, and avoidance of clearly harmful recommendations. These should be grounded in applicable clinical standards and established with input from qualified specialists.
Comparative performance could then be assessed against clinicians or human-AI teams appropriate to the intended workflow.
The relative contributions of generalist and specialist physicians deserve particular consideration.
We therefore recommend involving both groups in evaluation design while assigning responsibility for defining particular clinical acceptance criteria to professionals with qualifications appropriate to the clinical question.
The degree of human involvement in analogous real-world operation should also determine which comparator is most relevant. If a system is intended to support a clinician who reviews and acts on its recommendations, the combined performance of the clinician and AI may be a central endpoint. If the system makes decisions or takes actions autonomously, its independent performance may be assessed.
Patient-facing systems present an additional consideration. The absence of professional review at the point of interaction may necessitate more conservative safety requirements, stronger escalation pathways, and ongoing monitoring. However, retrospective human review should not automatically be treated as equivalent to clinician supervision before a consequential response reaches a patient.
Finally, because generative systems can produce many different clinically acceptable responses, scoring should focus on whether the response satisfies clinically meaningful criteria rather than whether it matches a single reference answer.
Question 15: Real-World Alternatives
Should AI medical devices be evaluated against the care patients would realistically receive without them, including delayed care or no treatment?
We strongly support consideration of the care or intervention that would realistically occur in the absence of a generative AI-enabled device. This is particularly important for applications intended to improve access-to-care.
In behavioral health, for example, the relevant alternative to an AI-enabled support tool may not always be an appointment with a licensed therapist. For some individuals, the alternative may be delayed care, limited access to clinical information, reliance on informal support, or no intervention at all.
Evaluating a device exclusively against an idealized healthcare scenario may therefore underestimate its potential benefits, particularly for populations facing substantial barriers to access.
However, we recommend distinguishing benefit relative to the available alternative from the device's absolute safety. A device may offer meaningful benefits where no timely care would otherwise be available while still creating risks that require independent assessment.
One particularly important question is whether an AI interaction introduces a clinically significant harm that would not otherwise have occurred.
For example, in behavioral health, a user experiencing ordinary sadness may seek conversational support. An AI system that amplifies emotional distress, or encourages inappropriate dependence could potentially worsen the user's condition, even if it never provides an explicitly dangerous instruction.
Conversely, a well-designed system may provide helpful information, encourage appropriate care-seeking, or offer beneficial support during periods when professional care is unavailable.
We recommend that clinical confirmation consider whether the device produces a net benefit, neutral effect, or additional harm relative to a realistic counterfactual, while separately measuring serious adverse outcomes that should not be offset by average benefit.
Potential comparators could include usual care, delayed specialist review, unaided clinical judgment, available digital resources, or no intervention, depending on the intended use and actual deployment setting.
For devices performing clinical reasoning or decision-making, comparison with qualified clinicians remains important. Specialist judgment may be particularly relevant for assessing complex cases, while generalist judgment can help characterize the decisions likely to occur earlier in a patient's care pathway.
When feasible, studies could examine not only whether the device's decision agrees with clinical experts, but also whether exposure to the device changes patient understanding, decision-making, care-seeking behavior, or clinical outcomes.
Such assessments should account for baseline patient characteristics and other factors that may influence outcomes independently of the AI system.
We believe this approach would allow FDA to recognize the substantial potential of generative AI to expand access to healthcare without overlooking the possibility that some applications create new harms rather than simply performing existing clinical functions less effectively.
Question 16: Independent Evaluation
What role should qualified independent third parties play in AI medical device evaluation, and how can their participation preserve competition and innovation?
We strongly support the involvement of qualified independent third parties throughout the competency-based evaluation process. Blind spots are natural, and independent scrutiny is particularly valuable for identifying failure modes that may not be apparent to the organizations developing a device.
Independent evaluators can contribute during premarket development by identifying vulnerabilities, assessing clinically significant behaviors, and providing evidence that informs design decisions and safeguards. This need not be limited to a final assessment immediately before submission. Early and iterative evaluation can help manufacturers address risks while products are still being developed, rather than discovering them after deployment.
The resulting evidence can also establish a baseline for postmarket monitoring. By evaluating devices against consistent clinical factors over time, independent organizations can help manufacturers identify emerging vulnerabilities, quantify regressions, and determine whether modifications have affected previously demonstrated competencies.
We recommend that qualifications reflect both clinical and technical expertise. Evaluators should demonstrate the ability to define clinically meaningful failure categories, incorporate perspectives from relevant specialists and patient populations, conduct sufficiently broad and high-volume testing, and produce reproducible, auditable findings. Technical testing capacity alone should not be sufficient; evaluation methods must also demonstrate clinical validity.
Importantly, independent evaluation need not require access to a manufacturer's proprietary models, training data, system prompts, or internal architecture. For example, at Circuit Breaker Labs, we conduct external evaluations through API endpoint-based access to deployed systems, allowing manufacturers to retain control of their underlying intellectual property. This approach can reduce barriers to participation while enabling independent assessment of observable device behavior.
Finally, we suggest that CDRH avoid creating an evaluation framework that inadvertently concentrates market access among a small number of organizations. Qualified for-profit companies, nonprofit laboratories, academic groups, and other independent evaluators should be able to participate under transparent, technology-neutral qualification criteria.
Independence should be defined by the integrity of the evaluation process, not the legal or commercial structure of the organization conducting it. Such an approach would encourage competition in evaluation methodologies, encourage innovation in the evaluation process itself, protect manufacturer intellectual property, and support the development of safety infrastructure capable of keeping pace with innovation.
Question 17: Multimodal AI
How should competency-based evaluation adapt to medical AI systems that process voice, images, and other modalities beyond text?
We believe the competency-based approach can be applied across different model architectures, provided that the competencies are defined according to the device's intended clinical function rather than the particular architecture of its underlying model.
The central regulatory question should remain whether the final deployed device performs its intended function safely and effectively under foreseeable conditions. However, multimodal devices introduce additional failure modes that may not be adequately captured through evaluations designed primarily for text-based systems.
For example, a voice-enabled device may respond appropriately to an accurate transcript while failing to recognize clinically significant information when it is spoken with an unfamiliar accent, interrupted, or affected by background noise. We have observed this with our voice red-teaming offering at Circuit Breaker Labs. Similarly, an image-based system may accurately interpret a clear clinical image but behave differently when the image is incomplete, poorly illuminated, or inconsistent with accompanying patient information.
We therefore recommend evaluating both individual components and the complete end-to-end clinical interaction. Component-level testing can help identify whether a failure originated in speech recognition, image interpretation, retrieval, reasoning, or another process. However, successful performance by individual components does not necessarily establish that the complete device will behave safely when those components interact.
An especially important competency is the handling of conflicting information across modalities. For example, a patient's verbal description of a condition may conflict with an image, a physiological measurement, or a previously documented clinical history. Evaluation should establish whether the device appropriately recognizes uncertainty, gathers additional information, or defers when necessary rather than arbitrarily prioritizing one input source.
Multimodal evaluation should also consider patient-specific communication patterns, including accents, dialects, age-related language differences, interrupted speech, and changes in how users present information across the course of an interaction.
Multi-participant red-teaming, an offering Circuit Breaker Labs, is also critical to consider as some AI-enabled medical devices may involve one or more patients and clinicians interfacing with an AI product, where risk is between human-AI and human-human.
For generative models, additional considerations may include the accuracy of modeled state transitions, the system's response to uncertainty, and the consequences of incorrect predictions over multiple steps. A plausible simulated future state does not necessarily establish that the predicted outcome is clinically reliable.
These systems should be assessed through realistic scenarios in which initial information may be incomplete, assumptions may be incorrect, and errors can accumulate over time.
Postmarket Monitoring
Question 18: Premarket vs. Postmarket Evidence
Under what circumstances should the FDA accept greater uncertainty before approval in exchange for stronger postmarket monitoring?
We recognize the rationale for considering greater reliance on postmarket monitoring, particularly because generative AI-enabled medical devices may operate in settings where real-world usage patterns and the full distribution of clinical risks are not yet well characterized.
However, postmarket monitoring should complement a meaningful premarket evidence baseline. This is particularly important for emerging patient-facing applications. In behavioral health, for example, available real-world data may not adequately represent the range of ways patients discuss sensitive conditions. Privacy constraints, limited data availability, and the evolving nature of human-AI interactions can make it difficult to construct comprehensive evaluation datasets from historical conversations alone.
We therefore recommend that manufacturers use clinically-informed synthetic scenarios to supplement available real-world evidence during premarket evaluation. These scenarios can systematically explore clinically plausible variations in patient characteristics, language, symptom presentation, comorbidities, and conversational trajectories.
High-volume testing is especially valuable in this context because it enables manufacturers to examine a broad range of foreseeable failures before users encounter them. Clinical experts should be involved in defining the risks being tested and evaluating whether synthetic scenarios reflect meaningful clinical situations.
Postmarket monitoring can then assess whether performance remains consistent with that baseline, whether changes introduce new vulnerabilities, and whether previously unidentified failure modes emerge as real-world evidence accumulates.
Greater reliance on postmarket evidence may be reasonable when the device has demonstrated adequate initial safety, the remaining uncertainties can be monitored meaningfully, and mechanisms exist to identify and remediate clinically significant problems in a timely manner.
For patient-facing conversational systems, human psychology introduces further considerations. A device's outputs may influence trust, emotional reliance, care-seeking behavior, and subsequent disclosures. These effects may develop across extended interactions and should not be assumed to be captured by conventional single-turn benchmarks.
We therefore recommend an approach in which robust premarket synthetic and clinical evaluation establishes a safety baseline, while continuous monitoring expands the evidence base and detects deviations after deployment.
This would support innovation without treating patients as the primary means of discovering risks that could reasonably have been identified before market entry. Humans should not be Guinea pigs.
Question 19: Postmarket Monitoring
How frequently should medical AI devices be reassessed, and what changes or events should trigger additional safety testing?
We support periodic re-benchmarking, sample-based clinician review, and performance degradation monitoring. However, we recommend that CDRH explicitly distinguish event-triggered reassessment from time-based reassessment, and consider both essential components of postmarket oversight.
Event-triggered evaluations are valuable when manufacturers make known changes that may affect device performance. Examples include modifications to the underlying foundation model, prompts, retrieval architecture, safety controls, guardrails, or clinical functionality.
However, manufacturers may not always know when their systems have changed in clinically significant ways. Third-party model providers may modify hosted services, previously unrecognized vulnerabilities may emerge, and changes in deployment conditions may affect behavior without a clearly identifiable software modification.
For this reason, postmarket monitoring should not depend exclusively on manufacturers recognizing that a change has occurred.
We recommend recurring re-benchmarking against established clinical safety criteria, even when no known modification has taken place. Such evaluations could identify clinically meaningful regressions, newly emerging vulnerabilities, or changes in the consistency of safety-critical behavior.
The appropriate cadence should be proportionate to the device's risk profile, anticipated usage, clinical setting, and rate of technical change. Monthly or quarterly evaluation may be appropriate for some applications, while higher-risk functions may require more frequent monitoring.
Clinical expertise should remain central throughout this process. Clinicians should contribute to defining evaluation criteria, identifying emerging clinical risks, and adjudicating cases where automated methods cannot reliably determine the clinical significance of an output.
We further recommend that monitoring include refreshed or independently developed challenge cases to help detect vulnerabilities that may not appear in previously used benchmarks.
The objective should be to create a monitoring system that identifies both regressions in known competencies and failures that were not anticipated during initial evaluation.
Question 20: Automated AI Oversight
How should the FDA evaluate the accuracy, reliability, and limitations of automated systems used to monitor other medical AI devices?
Machine-based supervisory systems can absolutely play an important role in scalable postmarket monitoring, particularly where devices produce large volumes of open-ended outputs and tests are necessary for generative AI systems, but cannot feasibly be reviewed individually by clinicians.
However, we caution against assuming that one generative model can reliably validate another. LLM-as-a-judge approaches exhibit several documented limitations, including:
These concerns are especially significant when supervisory systems are used to determine whether a safety-critical behavior has improved, degraded, or remained unchanged.
A broader concern is correlated error. When the device under evaluation and the supervisory model share similar assumptions, training influences, or failure modes, agreement between the two may create a false impression of independent validation.
We therefore recommend that CDRH encourage the use of complementary and independently validated evaluation mechanisms, including non-generative methods, specialized scoring systems, and qualified clinician review where appropriate.
Non-generative evaluation techniques can be designed to produce reproducible numerical scores while assessing clinically relevant contextual features. Circuit Breaker Labs, for example, develops clinically-informed machine-based evaluation methods that do not rely on a general-purpose LLM to assign final judgments.
We recommend that manufacturers characterize the reliability of any supervisory evaluation system by requesting repeated assessments of identical outputs.
CDRH should also consider evaluation throughput and operational scalability. A central advantage of machine-based supervision is the ability to evaluate substantially larger volumes of interactions than manual review alone. However, this advantage is meaningful only when the evaluation method is sufficiently accurate, reliable, and clinically appropriate.
Ultimately, the supervisory system should itself be treated as a measurement instrument whose validity, reliability, and limitations must be demonstrated. A changing device cannot be meaningfully monitored using an evaluation method whose own behavior is insufficiently characterized.
Question 21: Stakeholder Accountability
How should clinicians, healthcare institutions, independent evaluators, and other stakeholders contribute to postmarket monitoring while keeping manufacturers accountable?
All these stakeholders have valuable roles in postmarket monitoring, but CDRH should also recognize the important contributions of independent, for-profit evaluation organizations. Given the pace at which generative AI is advancing, a sustainable safety ecosystem must be capable of innovating at a comparable pace to for-profit manufacturers. Commercial evaluation organizations can have incentives to invest in scalable testing infrastructure, develop new methodologies, and rapidly adapt to emerging risks and new capabilities. Their involvement should be encouraged alongside academic institutions, professional societies, clinicians, and other stakeholders.
Shared participation in monitoring need not mean shared accountability for remediation. Manufacturers can remain responsible for defining their device's risk profile and demonstrating that appropriate safeguards are in place. Independent evaluators and other stakeholders can supplement this work by identifying previously unrecognized vulnerabilities, assessing performance against established clinical criteria, and detecting emerging risks through postmarket monitoring and reporting those back to manufacturers.
When monitoring identifies a clinically significant issue, the manufacturer should be responsible for investigating and addressing it within a timeframe proportionate to its severity. Where remediation cannot be completed within that timeframe, manufacturers should document a corrective action plan, interim risk mitigations, and an expected resolution timeline, with reporting to FDA where applicable. Particularly serious risks may require immediate intervention rather than awaiting completion of a remediation plan.
This approach allows different stakeholders to contribute their expertise without obscuring responsibility. Independent organizations can identify and characterize failures, clinicians and professional societies can inform their clinical significance, and manufacturers remain accountable for maintaining the safety and effectiveness of their devices.
Question 22: Re-Benchmarking After Changes
How should the extent of safety testing and regulatory review depend on the type and potential consequences of modifications to an AI device?
We recommend that post-deployment reassessment be proportional to the potential clinical consequences of a modification rather than solely its technical complexity or the manufacturer's intended purpose for the change.
A relatively minor software modification may have outsized implications for device safety. For example, changes to a system prompt, knowledge retrieval mechanism, context management strategy, or underlying foundation model may alter how a device interprets contraindications, recognizes clinical emergencies, or responds to patients outside its intended scope.
We propose a risk-based re-benchmarking framework built on the premarket competency assessment.
Manufacturers should maintain a mapping between their device's components, intended functions, clinically-significant failure categories, and established acceptance criteria. When a modification occurs, this mapping can help identify the competencies potentially affected and the evidence needed to demonstrate continued performance.
For example, a change to the user interface that does not materially affect clinical information or user decision-making may warrant limited reassessment. A change to the underlying language model, safety architecture, or conversational memory system may warrant broader testing, including multi-turn scenarios and relevant clinical subpopulations.
However, even targeted reassessment should be supplemented by cadenced safety-critical regression tests. Generative systems can exhibit unexpected interactions between components, and manufacturers should not assume that a modification affects only the functions they intended to change.
We recommend distinguishing regression testing from discovery-oriented testing. Regression testing determines whether established competencies have been maintained. Discovery-oriented testing uses refreshed, independently developed, or adversarial scenarios to identify new vulnerabilities that may emerge unexpectedly.
Changes that fall within a prospectively defined and appropriately validated modification framework may be suitable for management through a PCCP, consistent with applicable regulatory requirements. Changes that introduce new intended functions, substantially alter risk exposure, or exceed the boundaries of an established PCCP may require additional evidence or FDA review.
Importantly, the amount of evidence should be informed by the severity and detectability of potential harm, the degree of clinical autonomy, and the availability of effective downstream safeguards.
This approach would make re-benchmarking more efficient and predictable while recognizing that the safety impact of a modification cannot always be inferred from the size or apparent significance of the underlying engineering change.
Question 23: Unpredictable Modifications
How should predetermined change control plans (PCCPs) account for AI system changes that cannot be fully anticipated in advance?
PCCPs are valuable for modifications manufacturers can reasonably anticipate, including changes to underlying models, prompts, retrieval systems, or other components that may affect device performance. However, generative AI-enabled devices present an additional challenge: clinically meaningful behavioral changes may occur even when the manufacturer has not knowingly implemented a modification.
For example, a third-party foundation model provider may update a hosted model without providing sufficient visibility into the change. Even when the model identifier remains unchanged, differences in backend configuration or other dependencies may potentially affect performance.
We therefore recommend complementing modification-triggered PCCPs with proactive, recurring evaluation. Rather than testing only when a manufacturer anticipates that something has changed, a device could be periodically reassessed against its established clinical competencies and safety requirements, regardless of whether a modification has been identified.
This would create two complementary safeguards: predefined change-control procedures for anticipated modifications, and ongoing monitoring to detect unexpected changes, previously unidentified vulnerabilities, or performance degradation.
Importantly, the monitoring framework should define which safety-critical behaviors must remain consistent, how changes will be measured, what constitutes a clinically significant regression, and what corrective actions would follow. Evaluation methods should produce numerical, sufficiently reproducible results that can be tracked to distinguish changes in device performance from variability in the measurement process itself.
This approach would allow manufacturers to prepare for the reality that not every future change can be predicted, without requiring them to prespecify every possible modification. The central principle is that manufacturers may not know what will change, but they can define what safety-critical behaviors must continue to hold.
Question 24: Third-Party Model Updates
How can medical device manufacturers detect and address safety risks introduced by updates to third-party foundation models?
Third-party foundation model dependencies create a significant challenge because manufacturers may not control, anticipate, or even be notified of changes that affect their device's clinical performance.
We recommend that manufacturers implement ongoing behavioral monitoring and comparative evaluations of their deployed systems. Independent evaluation organizations, including companies such as Circuit Breaker Labs, can support this process through recurring safety assessments and controlled comparisons, or bakeoffs, between existing and prospective model configurations. For example, when a manufacturer considers replacing an underlying foundation model, the current and proposed versions of the complete device can be evaluated against common clinical safety criteria, including relevant failure modes and population-specific requirements. Because Circuit Breaker Labs produces numerical, explainable results, manufacturers can identify potential improvements and regressions before implementing the change.
The same approach can help detect unexpected changes introduced by third-party model providers. By periodically evaluating the deployed device against an established baseline, manufacturers can identify behavioral regressions even when they have not received notification of an underlying modification.
We believe recurring evaluation should be a reasonable expectation for GenAI-enabled medical devices, with frequency proportionate to clinical risk and the pace of system change. Monthly or quarterly assessments may be appropriate for some applications, while high-consequence systems or those dependent on frequently updated foundation models may warrant more frequent or event-triggered testing.
Importantly, periodic testing should complement, rather than replace, evaluations conducted before known model changes are deployed. Manufacturers should also consider technical and contractual mechanisms such as model-version pinning, advance update notifications, access to previous versions, and rollback capabilities where available.
Advances in automated, clinically-informed evaluation make broad regression testing substantially more scalable than exclusively manual approaches. This creates an opportunity to maintain rigorous oversight without imposing the same cost and operational burden as repeated full-scale clinical studies.
Ultimately, the appropriate regulatory expectation should not be that manufacturers can prevent every third-party model change. It should be that they maintain reliable mechanisms to detect clinically significant changes, assess their consequences, and take timely corrective action, regardless of whether the change originated within or outside their organization.
Other Topics
Question 25: Foundation Model Documentation
Are voluntary Foundation Model Master Files (MAFs) sufficient to support regulatory review, or should the FDA consider alternative evidence mechanisms?
We appreciate the potential value of voluntary Foundation MAFs, particularly as a mechanism for improving transparency about underlying models used in medical devices. However, we are concerned that voluntary disclosure alone may not provide sufficiently current or application-specific evidence of safety.
Foundation model developers may be slow to continually update regulatory documentation, particularly when they are not themselves developing the downstream medical device and may not be attuned to the specific risks.
A more fundamental limitation is that foundation model safety is not equivalent to device safety. A foundation model's performance depends substantially on how it is incorporated into a specific application. System prompts, retrieval-augmented generation, conversational memory, and guardrails can all introduce new vulnerabilities or alter existing ones.
Similarly, the clinical safety requirements for a behavioral health application may differ substantially from those for a medication management system, even when both use the same underlying foundation model.
For these reasons, we would not recommend relying on a voluntary MAF as the primary mechanism for establishing or maintaining the safety of downstream devices.
Instead, CDRH could consider a complementary framework involving ongoing independent foundation model evaluations and application-specific testing.
For example, qualified independent evaluation organizations could periodically compare major foundation models against common healthcare-relevant competencies, including clinical knowledge, safety-critical recognition, uncertainty communication, and consistency of behavior.
Such comparative evaluations, “bakeoffs” as Circuit Breaker Labs refers to them, could provide useful evidence about the relative strengths, limitations, and changing performance of different underlying models.
CDRH could consider maintaining or referencing a body of independently-generated evaluation evidence, with clear documentation of model versions, assessment methods, intended contexts, and known limitations. This could provide reviewers with additional information when assessing devices built on commonly used foundation models.
At Circuit Breaker Labs, comparative evaluation of different model configurations is one approach we use to help characterize clinically relevant performance differences. We believe a similar methodology could contribute to a broader evidence ecosystem without requiring foundation model developers to disclose proprietary training information or pending non-public updates.
However, these comparisons should not be interpreted as general authorization or certification of an underlying model for medical use. A model that performs well on common healthcare benchmarks may still behave unsafely when integrated into a particular device or altered via RAG, etc.
If this approach is taken, we recommend treating foundation model evidence as a useful reference that can inform, but not replace, application-specific evaluation.
Voluntary MAFs could still contribute important information, including version identification, known limitations, major safety-relevant changes, and update-notification practices. These disclosures would be most useful when combined with proactive, recurring behavioral testing rather than relied upon as a static representation of model safety.
The objective should be to establish a regulatory evidence framework that remains informative even when model providers make unanticipated changes, use cases evolve, or the performance characteristics of a previously evaluated model shift over time.
Question 26: Agentic AI
What additional safety requirements and oversight are necessary for autonomous medical AI systems that perform multi-step actions and use external tools?
Agentic generative AI-enabled devices warrant additional scrutiny because their risks may emerge across sequences of decisions and actions rather than from any individual output.
This is particularly important for systems capable of using external tools, maintaining conversational context, modifying records, initiating clinical workflows, or performing actions with limited opportunities for human review.
We recommend that CDRH place particular emphasis on high-volume, clinically-informed, trajectory-level evaluation in both premarket and postmarket settings.
Generative AI systems can produce variable outputs in response to similar inputs, and agentic systems introduce further complexity because small differences in an early decision may propagate through subsequent actions. Evaluating a small number of isolated examples is therefore unlikely to adequately characterize the range of potential failures.
Manufacturers should evaluate realistic multi-step scenarios, including variations in the order of information, incomplete patient histories, conflicting inputs, changing clinical conditions, and extended interactions in which important safety information may become less salient.
A central concern is context degradation. An agent may initially recognize a clinically significant contraindication but subsequently fail to account for it when performing an action many steps later. Likewise, an agent may correctly execute individual tool calls while producing an unsafe overall outcome because it has lost track of the clinical context or failed to reassess whether its planned actions remain appropriate.
These risks are not hypothetical properties of tool use alone. They arise from the interaction between the underlying generative system, its context-management mechanisms, the tools it can access, and the clinical consequences of its decisions.
Our experience conducting trajectory-based evaluations in behavioral health has reinforced the importance of assessing complete interactions. The order in which information is introduced, the progression of the user's circumstances, and the accumulated conversational context can materially affect how safety safeguards are invoked and whether a system maintains appropriate boundaries.
We recommend that acceptance criteria therefore address not only the correctness of individual responses or tool calls, but also the safety of the complete action sequence.
Relevant competencies should include persistent recognition of clinically significant information, adherence to authorized scope, appropriate human-approval checkpoints, resistance to prompt injection, handling of erroneous tool outputs, and the ability to stop or safely recover when circumstances change.
Equally important is the scientific reliability of the evaluation process itself. Agentic systems produce complex, open-ended trajectories, but this does not mean their performance must be assessed through subjective or irreproducible judgments.
Evaluation methods can incorporate clinical context and nuanced behavioral criteria while producing structured, quantitative, and auditable results. We recommend that manufacturers demonstrate how their scoring methods distinguish clinically significant failures, how reliably those methods agree with qualified clinical judgment, and whether repeated evaluation of identical trajectories produces consistent results.
A zero-variance evaluator should assign the same score to an identical trajectory. Such reproducibility is particularly valuable for identifying safety regressions across device versions and postmarket updates. Generative LLM-as-a-judge approaches may provide useful qualitative assessments, but their variability, systematic biases, and potential correlated errors should be characterized before they are relied upon for safety-critical determinations. A changing evaluation method can make it difficult to determine whether a device has actually improved or deteriorated.
Finally, we recommend that agentic device evaluation combine large-scale synthetic testing with clinically representative confirmation and continuing postmarket reassessment. Testing should be sufficiently broad to explore complex and uncommon trajectories, while statistical claims should account for the representativeness and dependence of generated scenarios.
The objective should be to establish reproducible evidence that clinically significant safeguards persist across complex action sequences, changing contexts, and evolving system configurations.
As generative AI becomes increasingly capable of acting rather than merely providing information, the importance of measuring these properties will only grow.
FDA Comment
Considerations for the Regulation of Generative AI-Enabled Medical Devices
Circuit Breaker Labs | FDA Public Comment • October 19, 2026
Assessing Risk
Question 1: Risk Framework
Does the proposed two-axis risk framework adequately capture the risks of generative AI medical devices, or should additional factors be considered?
The proposed two-axis framework provides a useful foundation, particularly because it recognizes risk as a continuum. However, device activity and the consequences of an incorrect output do not fully characterize the risks of generative AI-enabled medical devices. We recommend retaining the two axes while introducing additional axes.
First, risk depends on the characteristics of the intended user. A patient, a practicing clinician, a specialist, and a clinician in training may interact with the same system differently and have different abilities to recognize an incorrect response. Even within these populations, age, health literacy, clinical complexity, and patterns of use can materially affect risk. For instance, adolescent users may use language that differs substantially from the system's expected inputs. Risk assessment should consider foreseeable use behaviors rather than assuming all users interact with the device as intended.
Second, informational functions warrant greater differentiation. A resource navigator that directs users to established clinical information differs meaningfully from a system that synthesizes medication information in the context of an individual's treatment history. Even without an explicit recommendation, personalized information may influence clinical decisions. The extent to which a system incorporates patient-specific information should therefore modify its risk classification.
Third, the presence of human supervision should not automatically be treated as equivalent to effective oversight. A clinician who reviews every output before it reaches a patient provides a different safeguard from one who retrospectively reviews a small sample of interactions. Risk assessment should consider the timing, frequency, coverage, clinical competence, and intervention authority associated with human oversight. Retrospective sampling may detect systematic problems but cannot prevent harm from an individual output already delivered.
Fourth, traceability and robustness should be considered together. A system may correctly handle a clinically significant concept when expressed using formal terminology but respond differently when the same information is expressed colloquially. For example, requests involving “acetaminophen” and “Tylenol” should not produce materially different safety-critical behavior merely because of the terminology used, but we have observed materially different behavior in our evaluations at Circuit Breaker Labs. Traceability to appropriate clinical sources is valuable, but it does not itself establish that the system will apply those sources consistently.
Rather than adding numerous independent axes, CDRH could establish a small set of risk modifiers and associated testing expectations. This would preserve the flexibility of the proposed framework while giving manufacturers more predictable, application-specific requirements.
Question 2: Informational Functions
How should the FDA assess the risks of AI-generated information based on its wording, personalization, context, and influence on user decisions?
The degree to which an informational function directs behavior cannot be assessed solely through explicit instructions or recommendations. Risk also depends on how the information is communicated, the user's circumstances, and the degree of authority or interpersonal trust the system may implicitly convey.
A particularly important consideration is anthropomorphism, the attribution of human characteristics or understanding to an AI system. A system that speaks in a highly personal, emotionally affirming, or companion-like manner may encourage users to place greater confidence in its outputs, even when those outputs are nominally non-directive. Generative systems may also mirror or reinforce a user's beliefs and emotions, a behavior that can be inappropriate in certain clinical contexts.
We recommend a principle of minimum necessary anthropomorphism: manufacturers should justify personifying design features according to their intended clinical benefit and assess whether those features introduce undue trust, emotional dependence, or inappropriate behavioral influence. This need not prohibit first-person language or empathetic communication. Indeed, some therapeutic applications may benefit from carefully designed relational interactions. The important distinction is between communication that supports appropriate engagement and communication that encourages users to overestimate the system's understanding, competence, or authority.
CDRH could provide greater predictability through a behavioral assessment framework considering the use of patient-specific information, implied authority, emotional influence, and persistence of advice across repeated interactions.
We also recommend that manufacturers evaluate these properties across relevant patient populations, clinical conditions, and conversational contexts. Attachment-related risks may emerge gradually, particularly during extended interactions, and may not be apparent from isolated responses.
Rather than requiring manufacturers to eliminate all personalization, CDRH should encourage empirical evaluation of how personalization affects comprehension, trust calibration, and downstream decision-making.
Question 3: Patient-Facing AI
How should the FDA account for the unique risks of delivering AI-generated clinical information directly to patients rather than clinicians?
Patient-facing generative AI applications offer substantial potential benefits, including more timely access to health information, improved patient understanding, and support when a clinician is not immediately available. These benefits should not be discounted merely because patients lack formal clinical training.
However, patient-facing systems may require stronger safeguards because the recipient may not be able to identify an incorrect output or recognize when an apparently routine question involves clinically significant risk.
We recommend an approach in which multiple complementary safeguards address different failure modes. These should include clinically-grounded constraints, appropriate uncertainty communication, reliable escalation pathways, and testing against foreseeable real-world inputs.
An important consideration is the reliability of the safeguards themselves. Some systems use generative language models to evaluate or moderate the outputs of other generative models (‘LLM-as-a-judge’). Such approaches can introduce additional model-dependent error or variability. For safety-critical judgments, manufacturers should characterize the evaluator's sensitivity, specificity, reproducibility, variance, and susceptibility to adversarial or out-of-distribution inputs.
Patient-facing systems should also be evaluated using realistic input variation, including misspellings, colloquialisms, incomplete medical histories, contradictory information, differences in health literacy, and population-specific language. The goal should be reliable performance across the intended population.
Clinical claims should be supported by appropriate, current evidence, with uncertainty clearly communicated where evidence is limited. In clinically ambiguous or high-consequence circumstances, systems should be able to defer or escalate appropriately without over-refusing routine informational requests.
Importantly, these safeguards should be assessed through observable behavior, not solely through documentation that a safety policy or guardrail has been implemented.
Question 4: Clinical Specialization
How should differences between generalist and specialist physicians affect the risk assessment and safeguards of clinician-facing AI?
The distinction between generalist and specialist clinicians is important, but risk assessment should reflect the actual clinical workflow. We recommend involving both generalist and specialist clinicians in the development and evaluation of relevant GenAI-enabled functions. Generalists can help assess how a system is likely to be used in routine practice, while specialists can identify subtle errors, contraindications, and clinically significant failure modes that might otherwise go undetected.
This is particularly important for systems that provide specialized information to users without corresponding specialist training. A physician may appropriately rely on an AI system for support outside their primary field but may be less equipped to recognize when its recommendations are subtly incorrect.
Behavioral health provides a useful example. An AI system that repeatedly reassures or validates a user's concerns may appear an empathetic sounding board. However, in the context of obsessive-compulsive disorder, repetitive reassurance can reinforce maladaptive cycles. Similarly, uncritical affirmation of delusional beliefs may be clinically inappropriate. Evaluations developed without relevant specialist input may fail to distinguish appropriate emotional support from harmful reinforcement.
We recommend that clinical specialists contribute not only to reviewing outputs, but also to defining failure categories, evaluation criteria, and clinically appropriate safeguards. Generalist clinicians should participate in realistic workflow testing to determine whether the system communicates these distinctions effectively to its actual users.
For applications whose intended use population could reasonably span multiple clinical domains, the evaluation should also include relevant adjacent specialties and common comorbid conditions. A device intended for one specialty may create risks outside that specialty through medication interactions, overlapping symptoms, or treatment decisions.
Question 5: Multi-Turn Conversations
How should the FDA evaluate risks that emerge or change over the course of extended AI conversations?
Multi-turn conversational devices introduce risks that cannot be adequately characterized through isolated prompts. Safety-critical behavior may change over the course of a conversation as information accumulates, earlier details become less salient, and the user's presentation evolves.
One particularly important concern is context pollution. Clinically consequential information introduced early, e.g., during patient history, may no longer influence later outputs appropriately because of context-window limitations, summarization errors, retrieval failures, or the model's inconsistent use of available information.
For example, a patient may disclose a serious medication allergy early in an interaction, then return to medication-related questions much later. The continued absence of an allergic reaction does not make that contraindication irrelevant. Similarly, a system instructed to provide non-directive information may gradually transition toward individualized treatment recommendations as conversation context accumulates.
We recommend that CDRH encourage evaluation of complete conversational trajectories, including scenarios in which risks emerge gradually, relevant information is disclosed indirectly, users resist escalation, and important details are separated by substantial conversational distance.
Testing should examine whether safety-critical constraints persist despite changes in topic, user framing, and conversation length. Manufacturers should characterize the reliability of their memory mechanisms, context-management strategies, and safety controls under realistic and adversarial conditions.
Where appropriate, safety-critical information should be preserved in structured, persistent representations rather than relying exclusively on the generative model to recall and correctly prioritize it. Such representations should themselves be tested for accuracy, appropriate updates, and protection against conflicting or misleading information.
We further recommend distinguishing turn-level safety from trajectory-level safety. A system may produce individually plausible responses while its cumulative behavior becomes unsafe. The unit of evaluation must therefore include the conversation and its clinical consequences, not merely each response in isolation.
Question 6: Care Escalation
How should manufacturers measure and balance the risks of escalating care too frequently versus failing to escalate when necessary?
Under-escalation and over-escalation should be characterized separately because their consequences differ across clinical contexts and cannot always be reduced to a common error measure.
Repeated unnecessary escalation may also discourage patients from engaging with the system or seeking appropriate care in the future.
We recommend defining escalation thresholds according to the severity of potential harm, time sensitivity, uncertainty, availability of clinical supervision, and reversibility of an incorrect decision.
For instance, a clinician-supervised informational function may appropriately provide differentiated risk information without reflexively escalating every ambiguous presentation. Conversely, an autonomous function operating without timely human review may require a lower threshold for escalation in high-consequence situations.
Clinical prevalence and patient heterogeneity should also inform evaluation. In rare or highly variable conditions, the absence of a familiar symptom pattern may provide limited reassurance. However, rarity alone should not automatically mandate escalation. The relevant consideration is whether uncertainty and potential harm justify additional review.
CDRH should encourage manufacturers to report under-escalation and over-escalation as distinct outcomes, with clinically justified thresholds for each. Evaluations should include true emergencies, borderline presentations, and clearly non-urgent cases.
Particular attention should be given to escalation timing. A system that ultimately recommends emergency care may still fail if it first provides several rounds of inappropriate reassurance or delays the recommendation beyond a clinically meaningful point.
Competency-Based Premarket Evaluation
Question 7: Competency-Based Evaluation
Is a framework combining device benchmarking with clinical confirmation appropriate for evaluating generative AI medical devices?
We support the proposed competency-based approach, particularly the distinction between scalable non-clinical benchmarking and clinical confirmation under representative conditions.
That said, representative, high-volume testing of open-ended generative systems is not impractical. As a point of evidence: Circuit Breaker Labs has designed an evaluation system for monitoring of clinical AI that can run hundreds of thousands of dynamic, multi-turn tests each day, building meaningful clinical evidence at rapid speed.
Importantly, competency should not be understood as a static property established at the time of initial evaluation. Generative AI-enabled devices may change because of modifications to underlying models, prompts, guardrails, retrieval systems, user interfaces, or other components. Even changes not explicitly intended to affect clinical or safety performance may alter behavior in a clinically significant way.
We therefore recommend treating competency-based evaluation as the foundation of a continuous assurance process. Premarket benchmarking would establish a characterized baseline, clinical confirmation would provide evidence of performance in representative use, and postmarket reassessment would determine whether that competency is maintained as the system evolves.
The core competencies and acceptance criteria remain stable to support meaningful longitudinal comparisons. At the same time, individual test cases should be refreshed and expanded to account for emerging failure modes, changing clinical practices, and previously unidentified populations or conditions.
This would preserve standardization without allowing initial benchmark performance to become a substitute for continued evidence of safety and effectiveness.
Question 8: Risk-Based Evidence
How should a device's risk classification determine the amount and type of evidence required before FDA approval?
The two-axis framework could provide the foundation for determining the rigor and scope of premarket evidence, but the additional risk modifiers discussed in Question 1 should also inform what manufacturers are expected to demonstrate.
In particular, CDRH should consider the intended user, degree of personalization, clinical supervision model, patient population, potential for extended interactions, and clinical dependencies extending beyond the device's primary specialty.
A device providing general cardiovascular education, for example, should not necessarily face the same evidentiary expectations as one generating individualized cardiovascular treatment recommendations. Likewise, a cardiology-focused device may need to demonstrate appropriate handling of pulmonary disease, renal impairment, and other relevant comorbidities when they influence its intended function.
We recommend that manufacturers establish a risk-to-evidence mapping before testing. This would identify which clinically significant behaviors must be evaluated, which populations and deployment conditions must be represented, and what acceptance thresholds are justified for each category.
This mapping would also provide a reproducible starting point for postmarket evaluation. When device components change, manufacturers could determine which risk categories may have been affected and conduct targeted reassessment, while retaining broader regression testing for unexpected effects.
A standardized approach to mapping risk categories to evidence requirements could improve regulatory predictability without imposing identical testing obligations on devices with materially different risk profiles.
Question 9: Benchmarking Standards
Does the proposed benchmarking framework adequately assess clinical knowledge, reasoning, safety, communication, and generalizability, or are additional standards needed?
The proposed benchmarking elements provide a strong foundation. We particularly support the inclusion of safety-critical escalation, scope adherence, uncertainty communication, robustness, subgroup performance, and agentic capabilities.
However, we recommend some refinements.
First, CDRH should distinguish reusable evaluation criteria from reusable test instances. Standardized clinical competencies, scoring definitions, and acceptance criteria are essential for comparability. Repeated reliance on a fixed, publicly available set of prompts or data is less reliable, particularly when models or applications are often optimized against those exact examples. The evaluation framework should therefore support standardized evaluation metrics applied to both reference cases and independently generated, sequestered, or refreshed scenarios.
Second, clinically significant context retention should be an explicit competency. Devices should demonstrate that they maintain and appropriately apply relevant patient information across extended interactions, including contraindications, allergies, comorbidities, and changes in clinical status. This competency spans clinical reasoning, safety, robustness, and agentic behavior, but its importance warrants explicit consideration.
Third, anthropomorphism, inappropriate agreeableness, and automation bias should receive greater attention. A system may provide factually accurate information while encouraging excessive reliance or reinforcing harmful beliefs. These risks can emerge through communication style, emotional affirmation, and repeated interactions, particularly in behavioral health applications.
Fourth, clinical provenance should be assessed. Where a device presents information as clinically established, users should be able to distinguish established guidance from emerging evidence, inference, or uncertain claims. Unsupported assertions of authority should be considered separately from ordinary factual errors.
Finally, the proposed elements overlap in clinically meaningful ways. For example, an agent that reorders a medication without adequately considering a relevant contraindication may demonstrate failures of clinical proficiency, context handling, and agentic execution simultaneously.
We do not recommend treating these categories as mutually exclusive. Instead, CDRH could retain the organizing structure while allowing individual failures to map to multiple competencies, with careful safeguards against double-counting in aggregate scores.
Question 10: Benchmark Validity
How can manufacturers demonstrate that benchmark performance reliably predicts real-world safety without overfitting or introducing conflicts of interest?
CDRH identifies a central challenge in generative AI evaluation: performance on a known benchmark does not necessarily predict performance under novel, real-world conditions.
We recommend distinguishing the clinical construct being measured from the particular examples used to measure it. Clinical risk categories, evaluation criteria, and acceptance thresholds should be clearly specified and sufficiently stable to support reproducibility. However, the scenarios used to assess those should be capable of evolving as new failure modes and patterns of use emerge.
A robust approach could combine common reference benchmarks, independent sequestered cases, and dynamically generated scenarios tailored to a device's intended use. This would allow manufacturers to assess established competencies while reducing overreliance on a finite set of recognizable inputs.
The construct validity of such benchmarks should be supported through qualified clinical review, explicit links to clinically meaningful failure mechanisms, testing across realistic input variation, and comparison with independent clinical adjudication where appropriate.
We also recommend evaluating measurement reliability separately from model performance. When a scoring method produces inconsistent judgments across equivalent inputs or repeated assessments, an apparent change in device performance may reflect measurement variation rather than an actual change in safety. Manufacturers should characterize the repeatability, agreement with qualified adjudicators, and limitations of their evaluation methods.
Independent evaluation organizations can help identify blind spots arising from sponsor-developed test suites, particularly when they collaborate with clinical specialists. However, independence should not preclude device-specific customization. A useful structure could include a common set of clinically defined risks supplemented by independently developed scenarios addressing the manufacturer's intended use, population, and known vulnerabilities.
Manufacturers should document what risks their systems may encounter, why those risks matter, and independent evaluators can assess handling of these risks and provide detail on how the scenarios represent intended use, and what important uncertainties remain.
Finally, benchmark maintenance should be treated as a continuing scientific activity. As underlying models and clinical applications evolve, evaluation methods must be updated without sacrificing reproducibility or longitudinal comparability.
Question 11: Clinical Confirmation
How should manufacturers select appropriate clinical validation methods based on device risk, intended use, and real-world patient interactions?
We support a risk-proportionate approach to clinical confirmation that allows manufacturers to select methods appropriate to their device's intended use, degree of autonomy, and potential consequences of failure. However, we recommend that the selection process begin with a structured characterization of clinical risks rather than a determination of which study design is most convenient.
Manufacturers could first describe their device's risk profile using the proposed two-axis framework, supplemented by the additional considerations discussed in our earlier responses. This assessment should identify both the situations the device is intended to address and foreseeable circumstances in which its outputs could cause harm.
For example, a cardiology-focused system may not explicitly address pulmonary disease, yet many patients presenting with cardiovascular concerns also have respiratory conditions or overlapping symptoms. A risk assessment that excludes these clinical intersections could overlook consequential failure modes. CDRH could request additional evidence in areas where the intended function creates foreseeable dependencies on adjacent clinical specialties.
We recommend that manufacturers document these anticipated risks and provide corresponding performance evidence, including findings from qualified independent evaluators where appropriate. Independent evaluators can help identify blind spots that may be overlooked when manufacturers develop their own evaluation criteria, while clinical specialists and generalist physicians can provide complementary perspectives on clinical complexity and actual patterns of use.
The anticipated distribution of real-world inputs should inform both test coverage and the amount of testing performed. A device expected to support large volumes of patient interactions should be evaluated across a sufficiently broad range of scenarios to characterize uncommon but clinically significant failures.
High-throughput automated testing makes it feasible to examine many more scenarios than traditional manual evaluation alone.
For perspective, our work at Circuit Breaker Labs has focused on scalable, clinically-informed testing methods capable of evaluating large numbers of interactions efficiently. Such methods can supplement clinical confirmation by systematically examining variation in patient characteristics, language, comorbidities, and conversational trajectories.
It is also important to distinguish reproducibility of the evaluation method from variability in the device being evaluated. Reproducible scoring methods can make longitudinal comparisons more reliable, but their clinical validity must still be established through appropriate expert agreement and other evidence.
Finally, we recommend a staged approach in which scalable benchmarking identifies areas of uncertainty that then inform more resource-intensive confirmation methods. For instance, failures involving complex comorbidities discovered through synthetic testing could guide targeted retrospective analysis, standardized patient interactions, or shadow deployment.
This would allow manufacturers to use clinical evidence more efficiently while directing greater scrutiny toward the conditions most likely to produce significant harm.
Question 12: Statistical Reliability
How should manufacturers establish statistically meaningful performance estimates when combining synthetic testing, real-world data, and clinical evidence?
Statistically meaningful performance measurement is particularly challenging for generative AI-enabled devices because outputs may vary across repeated interactions, clinically significant failures may be uncommon, and the real-world distribution of patient inputs may be incompletely characterized.
We recommend that CDRH encourage large-scale, risk-stratified evaluation supported by clinically-informed synthetic data, while clearly distinguishing performance under synthetic testing conditions from estimated real-world performance.
Real-world patient data is essential for understanding how individuals actually interact with medical AI, including the conditions they present with, their language patterns, and the frequency of different clinical scenarios. However, particularly for emerging conversational and patient-facing applications, available data may not adequately represent the full range of potential interactions.
Privacy-preserving analysis of real-world data can help characterize these patterns and inform the construction of synthetic test scenarios. Clinical experts can then assess whether the generated cases are medically plausible, reflect meaningful patterns of presentation, and cover relevant populations and conditions.
Importantly, synthetic data should not merely reproduce the observed distribution of existing patient interactions. Real-world datasets may systematically underrepresent rare conditions, underserved populations, atypical symptoms, or clinically important scenarios that have not yet been encountered. Synthetic testing creates an opportunity to intentionally examine such situations before they result in real-world harm.
Advances in automated evaluation have made testing at the scale of hundreds of thousands or millions of interactions technically feasible, as evidenced by the in-market tool developed by Circuit Breaker Labs. We believe this capability should be leveraged to characterize performance across a broad range of clinically meaningful scenarios, including variations in patient language, age, comorbidities, and conversational histories.
We recommend separating three components of performance evidence:
We also recommend reporting uncertainty intervals, clinically relevant subgroup performance, and failure severity rather than relying exclusively on a single aggregate score.
The objective should be to make statistically rigorous evaluation more accessible and scalable.
Question 13: Synthetic Data
Where is synthetic data most useful for evaluating medical AI, and how can manufacturers prevent it from reproducing existing model biases and blind spots?
We believe synthetic data can make substantial contributions across clinical domains, device functions, and patient populations, particularly where real-world data is insufficient to characterize the full range of foreseeable inputs.
Conversational and behavioral health applications illustrate both the need for synthetic data and the importance of developing it carefully. Unlike many structured clinical datasets, patient-facing conversations can include indirect disclosures, evolving symptoms, incomplete histories, informal language, and substantial variation in how individuals describe similar experiences. Two patients presenting with the same underlying clinical concern may use entirely different language, and relatively small wording differences may influence whether an AI system recognizes a safety-critical condition or responds appropriately.
For example, a patient describing significant psychological distress may not use formal diagnostic terminology or explicitly identify the condition they are experiencing. A system evaluated primarily on textbook descriptions may perform well on a benchmark while failing to recognize clinically meaningful variations encountered in practice.
Synthetic testing offers a way to systematically explore these differences. It can generate variations in wording, age-related language, clinical context, comorbidities, symptom progression, and conversation length while preserving a clinically specified underlying scenario.
However, the clinical risks being tested should not be done solely by the same class of generative models being evaluated (‘LLM-as-a-judge approaches’). Otherwise judging may reproduce overlapping assumptions or blind spots.
Clinical experts should establish the relevant risk categories and behavioral expectations. Synthetic scenarios can then be generated through complementary techniques, including structured simulation and steering. Finally, outputs should demonstrate reliability and clinical validity.
At Circuit Breaker Labs, our approach emphasizes clinically-informed risk criteria and reproducible evaluation methods rather than solely on one generative model to independently produce scenarios and another to judge the results. We believe this separation is an important design principle for developing trustworthy evaluation infrastructure.
The involvement of clinicians experienced with underrepresented populations is especially important. Available patient datasets may not fully reflect how individuals across age groups, dialects, health literacy levels, cultural contexts, and co-occurring conditions describe their concerns. Specialists and generalists can identify clinically meaningful variations that would otherwise be absent from test-generation processes.
We also recommend that synthetic datasets be continuously expanded and validated as new evidence emerges from clinical practice, postmarket monitoring, patient feedback, and independent evaluations.
The appropriate goal is therefore not to replace real-world evidence, but to extend the range of conditions under which safety can be examined before patients encounter those conditions themselves.
Question 14: Performance Comparators
What clinical standards, physician expertise, or human-AI performance benchmarks should determine whether an AI device performs acceptably?
We recommend that performance comparators and acceptance criteria be determined according to a device's intended use, level of autonomy, patient population, and clinical consequences of failure.
The proposed axis framework could provide a starting point for establishing baseline expectations. Manufacturers should define the functions within the device's intended scope, the behaviors that would constitute unacceptable deviations from that scope, and the clinical competencies required for safe operation.
Minimum requirements could include recognition of clinically significant contraindications, appropriate escalation in emergencies, adherence to the device's intended scope, and avoidance of clearly harmful recommendations. These should be grounded in applicable clinical standards and established with input from qualified specialists.
Comparative performance could then be assessed against clinicians or human-AI teams appropriate to the intended workflow.
The relative contributions of generalist and specialist physicians deserve particular consideration.
We therefore recommend involving both groups in evaluation design while assigning responsibility for defining particular clinical acceptance criteria to professionals with qualifications appropriate to the clinical question.
The degree of human involvement in analogous real-world operation should also determine which comparator is most relevant. If a system is intended to support a clinician who reviews and acts on its recommendations, the combined performance of the clinician and AI may be a central endpoint. If the system makes decisions or takes actions autonomously, its independent performance may be assessed.
Patient-facing systems present an additional consideration. The absence of professional review at the point of interaction may necessitate more conservative safety requirements, stronger escalation pathways, and ongoing monitoring. However, retrospective human review should not automatically be treated as equivalent to clinician supervision before a consequential response reaches a patient.
Finally, because generative systems can produce many different clinically acceptable responses, scoring should focus on whether the response satisfies clinically meaningful criteria rather than whether it matches a single reference answer.
Question 15: Real-World Alternatives
Should AI medical devices be evaluated against the care patients would realistically receive without them, including delayed care or no treatment?
We strongly support consideration of the care or intervention that would realistically occur in the absence of a generative AI-enabled device. This is particularly important for applications intended to improve access-to-care.
In behavioral health, for example, the relevant alternative to an AI-enabled support tool may not always be an appointment with a licensed therapist. For some individuals, the alternative may be delayed care, limited access to clinical information, reliance on informal support, or no intervention at all.
Evaluating a device exclusively against an idealized healthcare scenario may therefore underestimate its potential benefits, particularly for populations facing substantial barriers to access.
However, we recommend distinguishing benefit relative to the available alternative from the device's absolute safety. A device may offer meaningful benefits where no timely care would otherwise be available while still creating risks that require independent assessment.
One particularly important question is whether an AI interaction introduces a clinically significant harm that would not otherwise have occurred.
For example, in behavioral health, a user experiencing ordinary sadness may seek conversational support. An AI system that amplifies emotional distress, or encourages inappropriate dependence could potentially worsen the user's condition, even if it never provides an explicitly dangerous instruction.
Conversely, a well-designed system may provide helpful information, encourage appropriate care-seeking, or offer beneficial support during periods when professional care is unavailable.
We recommend that clinical confirmation consider whether the device produces a net benefit, neutral effect, or additional harm relative to a realistic counterfactual, while separately measuring serious adverse outcomes that should not be offset by average benefit.
Potential comparators could include usual care, delayed specialist review, unaided clinical judgment, available digital resources, or no intervention, depending on the intended use and actual deployment setting.
For devices performing clinical reasoning or decision-making, comparison with qualified clinicians remains important. Specialist judgment may be particularly relevant for assessing complex cases, while generalist judgment can help characterize the decisions likely to occur earlier in a patient's care pathway.
When feasible, studies could examine not only whether the device's decision agrees with clinical experts, but also whether exposure to the device changes patient understanding, decision-making, care-seeking behavior, or clinical outcomes.
Such assessments should account for baseline patient characteristics and other factors that may influence outcomes independently of the AI system.
We believe this approach would allow FDA to recognize the substantial potential of generative AI to expand access to healthcare without overlooking the possibility that some applications create new harms rather than simply performing existing clinical functions less effectively.
Question 16: Independent Evaluation
What role should qualified independent third parties play in AI medical device evaluation, and how can their participation preserve competition and innovation?
We strongly support the involvement of qualified independent third parties throughout the competency-based evaluation process. Blind spots are natural, and independent scrutiny is particularly valuable for identifying failure modes that may not be apparent to the organizations developing a device.
Independent evaluators can contribute during premarket development by identifying vulnerabilities, assessing clinically significant behaviors, and providing evidence that informs design decisions and safeguards. This need not be limited to a final assessment immediately before submission. Early and iterative evaluation can help manufacturers address risks while products are still being developed, rather than discovering them after deployment.
The resulting evidence can also establish a baseline for postmarket monitoring. By evaluating devices against consistent clinical factors over time, independent organizations can help manufacturers identify emerging vulnerabilities, quantify regressions, and determine whether modifications have affected previously demonstrated competencies.
We recommend that qualifications reflect both clinical and technical expertise. Evaluators should demonstrate the ability to define clinically meaningful failure categories, incorporate perspectives from relevant specialists and patient populations, conduct sufficiently broad and high-volume testing, and produce reproducible, auditable findings. Technical testing capacity alone should not be sufficient; evaluation methods must also demonstrate clinical validity.
Importantly, independent evaluation need not require access to a manufacturer's proprietary models, training data, system prompts, or internal architecture. For example, at Circuit Breaker Labs, we conduct external evaluations through API endpoint-based access to deployed systems, allowing manufacturers to retain control of their underlying intellectual property. This approach can reduce barriers to participation while enabling independent assessment of observable device behavior.
Finally, we suggest that CDRH avoid creating an evaluation framework that inadvertently concentrates market access among a small number of organizations. Qualified for-profit companies, nonprofit laboratories, academic groups, and other independent evaluators should be able to participate under transparent, technology-neutral qualification criteria.
Independence should be defined by the integrity of the evaluation process, not the legal or commercial structure of the organization conducting it. Such an approach would encourage competition in evaluation methodologies, encourage innovation in the evaluation process itself, protect manufacturer intellectual property, and support the development of safety infrastructure capable of keeping pace with innovation.
Question 17: Multimodal AI
How should competency-based evaluation adapt to medical AI systems that process voice, images, and other modalities beyond text?
We believe the competency-based approach can be applied across different model architectures, provided that the competencies are defined according to the device's intended clinical function rather than the particular architecture of its underlying model.
The central regulatory question should remain whether the final deployed device performs its intended function safely and effectively under foreseeable conditions. However, multimodal devices introduce additional failure modes that may not be adequately captured through evaluations designed primarily for text-based systems.
For example, a voice-enabled device may respond appropriately to an accurate transcript while failing to recognize clinically significant information when it is spoken with an unfamiliar accent, interrupted, or affected by background noise. We have observed this with our voice red-teaming offering at Circuit Breaker Labs. Similarly, an image-based system may accurately interpret a clear clinical image but behave differently when the image is incomplete, poorly illuminated, or inconsistent with accompanying patient information.
We therefore recommend evaluating both individual components and the complete end-to-end clinical interaction. Component-level testing can help identify whether a failure originated in speech recognition, image interpretation, retrieval, reasoning, or another process. However, successful performance by individual components does not necessarily establish that the complete device will behave safely when those components interact.
An especially important competency is the handling of conflicting information across modalities. For example, a patient's verbal description of a condition may conflict with an image, a physiological measurement, or a previously documented clinical history. Evaluation should establish whether the device appropriately recognizes uncertainty, gathers additional information, or defers when necessary rather than arbitrarily prioritizing one input source.
Multimodal evaluation should also consider patient-specific communication patterns, including accents, dialects, age-related language differences, interrupted speech, and changes in how users present information across the course of an interaction.
Multi-participant red-teaming, an offering Circuit Breaker Labs, is also critical to consider as some AI-enabled medical devices may involve one or more patients and clinicians interfacing with an AI product, where risk is between human-AI and human-human.
For generative models, additional considerations may include the accuracy of modeled state transitions, the system's response to uncertainty, and the consequences of incorrect predictions over multiple steps. A plausible simulated future state does not necessarily establish that the predicted outcome is clinically reliable.
These systems should be assessed through realistic scenarios in which initial information may be incomplete, assumptions may be incorrect, and errors can accumulate over time.
Postmarket Monitoring
Question 18: Premarket vs. Postmarket Evidence
Under what circumstances should the FDA accept greater uncertainty before approval in exchange for stronger postmarket monitoring?
We recognize the rationale for considering greater reliance on postmarket monitoring, particularly because generative AI-enabled medical devices may operate in settings where real-world usage patterns and the full distribution of clinical risks are not yet well characterized.
However, postmarket monitoring should complement a meaningful premarket evidence baseline. This is particularly important for emerging patient-facing applications. In behavioral health, for example, available real-world data may not adequately represent the range of ways patients discuss sensitive conditions. Privacy constraints, limited data availability, and the evolving nature of human-AI interactions can make it difficult to construct comprehensive evaluation datasets from historical conversations alone.
We therefore recommend that manufacturers use clinically-informed synthetic scenarios to supplement available real-world evidence during premarket evaluation. These scenarios can systematically explore clinically plausible variations in patient characteristics, language, symptom presentation, comorbidities, and conversational trajectories.
High-volume testing is especially valuable in this context because it enables manufacturers to examine a broad range of foreseeable failures before users encounter them. Clinical experts should be involved in defining the risks being tested and evaluating whether synthetic scenarios reflect meaningful clinical situations.
Postmarket monitoring can then assess whether performance remains consistent with that baseline, whether changes introduce new vulnerabilities, and whether previously unidentified failure modes emerge as real-world evidence accumulates.
Greater reliance on postmarket evidence may be reasonable when the device has demonstrated adequate initial safety, the remaining uncertainties can be monitored meaningfully, and mechanisms exist to identify and remediate clinically significant problems in a timely manner.
For patient-facing conversational systems, human psychology introduces further considerations. A device's outputs may influence trust, emotional reliance, care-seeking behavior, and subsequent disclosures. These effects may develop across extended interactions and should not be assumed to be captured by conventional single-turn benchmarks.
We therefore recommend an approach in which robust premarket synthetic and clinical evaluation establishes a safety baseline, while continuous monitoring expands the evidence base and detects deviations after deployment.
This would support innovation without treating patients as the primary means of discovering risks that could reasonably have been identified before market entry. Humans should not be Guinea pigs.
Question 19: Postmarket Monitoring
How frequently should medical AI devices be reassessed, and what changes or events should trigger additional safety testing?
We support periodic re-benchmarking, sample-based clinician review, and performance degradation monitoring. However, we recommend that CDRH explicitly distinguish event-triggered reassessment from time-based reassessment, and consider both essential components of postmarket oversight.
Event-triggered evaluations are valuable when manufacturers make known changes that may affect device performance. Examples include modifications to the underlying foundation model, prompts, retrieval architecture, safety controls, guardrails, or clinical functionality.
However, manufacturers may not always know when their systems have changed in clinically significant ways. Third-party model providers may modify hosted services, previously unrecognized vulnerabilities may emerge, and changes in deployment conditions may affect behavior without a clearly identifiable software modification.
For this reason, postmarket monitoring should not depend exclusively on manufacturers recognizing that a change has occurred.
We recommend recurring re-benchmarking against established clinical safety criteria, even when no known modification has taken place. Such evaluations could identify clinically meaningful regressions, newly emerging vulnerabilities, or changes in the consistency of safety-critical behavior.
The appropriate cadence should be proportionate to the device's risk profile, anticipated usage, clinical setting, and rate of technical change. Monthly or quarterly evaluation may be appropriate for some applications, while higher-risk functions may require more frequent monitoring.
Clinical expertise should remain central throughout this process. Clinicians should contribute to defining evaluation criteria, identifying emerging clinical risks, and adjudicating cases where automated methods cannot reliably determine the clinical significance of an output.
We further recommend that monitoring include refreshed or independently developed challenge cases to help detect vulnerabilities that may not appear in previously used benchmarks.
The objective should be to create a monitoring system that identifies both regressions in known competencies and failures that were not anticipated during initial evaluation.
Question 20: Automated AI Oversight
How should the FDA evaluate the accuracy, reliability, and limitations of automated systems used to monitor other medical AI devices?
Machine-based supervisory systems can absolutely play an important role in scalable postmarket monitoring, particularly where devices produce large volumes of open-ended outputs and tests are necessary for generative AI systems, but cannot feasibly be reviewed individually by clinicians.
However, we caution against assuming that one generative model can reliably validate another. LLM-as-a-judge approaches exhibit several documented limitations, including:
These concerns are especially significant when supervisory systems are used to determine whether a safety-critical behavior has improved, degraded, or remained unchanged.
A broader concern is correlated error. When the device under evaluation and the supervisory model share similar assumptions, training influences, or failure modes, agreement between the two may create a false impression of independent validation.
We therefore recommend that CDRH encourage the use of complementary and independently validated evaluation mechanisms, including non-generative methods, specialized scoring systems, and qualified clinician review where appropriate.
Non-generative evaluation techniques can be designed to produce reproducible numerical scores while assessing clinically relevant contextual features. Circuit Breaker Labs, for example, develops clinically-informed machine-based evaluation methods that do not rely on a general-purpose LLM to assign final judgments.
We recommend that manufacturers characterize the reliability of any supervisory evaluation system by requesting repeated assessments of identical outputs.
CDRH should also consider evaluation throughput and operational scalability. A central advantage of machine-based supervision is the ability to evaluate substantially larger volumes of interactions than manual review alone. However, this advantage is meaningful only when the evaluation method is sufficiently accurate, reliable, and clinically appropriate.
Ultimately, the supervisory system should itself be treated as a measurement instrument whose validity, reliability, and limitations must be demonstrated. A changing device cannot be meaningfully monitored using an evaluation method whose own behavior is insufficiently characterized.
Question 21: Stakeholder Accountability
How should clinicians, healthcare institutions, independent evaluators, and other stakeholders contribute to postmarket monitoring while keeping manufacturers accountable?
All these stakeholders have valuable roles in postmarket monitoring, but CDRH should also recognize the important contributions of independent, for-profit evaluation organizations. Given the pace at which generative AI is advancing, a sustainable safety ecosystem must be capable of innovating at a comparable pace to for-profit manufacturers. Commercial evaluation organizations can have incentives to invest in scalable testing infrastructure, develop new methodologies, and rapidly adapt to emerging risks and new capabilities. Their involvement should be encouraged alongside academic institutions, professional societies, clinicians, and other stakeholders.
Shared participation in monitoring need not mean shared accountability for remediation. Manufacturers can remain responsible for defining their device's risk profile and demonstrating that appropriate safeguards are in place. Independent evaluators and other stakeholders can supplement this work by identifying previously unrecognized vulnerabilities, assessing performance against established clinical criteria, and detecting emerging risks through postmarket monitoring and reporting those back to manufacturers.
When monitoring identifies a clinically significant issue, the manufacturer should be responsible for investigating and addressing it within a timeframe proportionate to its severity. Where remediation cannot be completed within that timeframe, manufacturers should document a corrective action plan, interim risk mitigations, and an expected resolution timeline, with reporting to FDA where applicable. Particularly serious risks may require immediate intervention rather than awaiting completion of a remediation plan.
This approach allows different stakeholders to contribute their expertise without obscuring responsibility. Independent organizations can identify and characterize failures, clinicians and professional societies can inform their clinical significance, and manufacturers remain accountable for maintaining the safety and effectiveness of their devices.
Question 22: Re-Benchmarking After Changes
How should the extent of safety testing and regulatory review depend on the type and potential consequences of modifications to an AI device?
We recommend that post-deployment reassessment be proportional to the potential clinical consequences of a modification rather than solely its technical complexity or the manufacturer's intended purpose for the change.
A relatively minor software modification may have outsized implications for device safety. For example, changes to a system prompt, knowledge retrieval mechanism, context management strategy, or underlying foundation model may alter how a device interprets contraindications, recognizes clinical emergencies, or responds to patients outside its intended scope.
We propose a risk-based re-benchmarking framework built on the premarket competency assessment.
Manufacturers should maintain a mapping between their device's components, intended functions, clinically-significant failure categories, and established acceptance criteria. When a modification occurs, this mapping can help identify the competencies potentially affected and the evidence needed to demonstrate continued performance.
For example, a change to the user interface that does not materially affect clinical information or user decision-making may warrant limited reassessment. A change to the underlying language model, safety architecture, or conversational memory system may warrant broader testing, including multi-turn scenarios and relevant clinical subpopulations.
However, even targeted reassessment should be supplemented by cadenced safety-critical regression tests. Generative systems can exhibit unexpected interactions between components, and manufacturers should not assume that a modification affects only the functions they intended to change.
We recommend distinguishing regression testing from discovery-oriented testing. Regression testing determines whether established competencies have been maintained. Discovery-oriented testing uses refreshed, independently developed, or adversarial scenarios to identify new vulnerabilities that may emerge unexpectedly.
Changes that fall within a prospectively defined and appropriately validated modification framework may be suitable for management through a PCCP, consistent with applicable regulatory requirements. Changes that introduce new intended functions, substantially alter risk exposure, or exceed the boundaries of an established PCCP may require additional evidence or FDA review.
Importantly, the amount of evidence should be informed by the severity and detectability of potential harm, the degree of clinical autonomy, and the availability of effective downstream safeguards.
This approach would make re-benchmarking more efficient and predictable while recognizing that the safety impact of a modification cannot always be inferred from the size or apparent significance of the underlying engineering change.
Question 23: Unpredictable Modifications
How should predetermined change control plans (PCCPs) account for AI system changes that cannot be fully anticipated in advance?
PCCPs are valuable for modifications manufacturers can reasonably anticipate, including changes to underlying models, prompts, retrieval systems, or other components that may affect device performance. However, generative AI-enabled devices present an additional challenge: clinically meaningful behavioral changes may occur even when the manufacturer has not knowingly implemented a modification.
For example, a third-party foundation model provider may update a hosted model without providing sufficient visibility into the change. Even when the model identifier remains unchanged, differences in backend configuration or other dependencies may potentially affect performance.
We therefore recommend complementing modification-triggered PCCPs with proactive, recurring evaluation. Rather than testing only when a manufacturer anticipates that something has changed, a device could be periodically reassessed against its established clinical competencies and safety requirements, regardless of whether a modification has been identified.
This would create two complementary safeguards: predefined change-control procedures for anticipated modifications, and ongoing monitoring to detect unexpected changes, previously unidentified vulnerabilities, or performance degradation.
Importantly, the monitoring framework should define which safety-critical behaviors must remain consistent, how changes will be measured, what constitutes a clinically significant regression, and what corrective actions would follow. Evaluation methods should produce numerical, sufficiently reproducible results that can be tracked to distinguish changes in device performance from variability in the measurement process itself.
This approach would allow manufacturers to prepare for the reality that not every future change can be predicted, without requiring them to prespecify every possible modification. The central principle is that manufacturers may not know what will change, but they can define what safety-critical behaviors must continue to hold.
Question 24: Third-Party Model Updates
How can medical device manufacturers detect and address safety risks introduced by updates to third-party foundation models?
Third-party foundation model dependencies create a significant challenge because manufacturers may not control, anticipate, or even be notified of changes that affect their device's clinical performance.
We recommend that manufacturers implement ongoing behavioral monitoring and comparative evaluations of their deployed systems. Independent evaluation organizations, including companies such as Circuit Breaker Labs, can support this process through recurring safety assessments and controlled comparisons, or bakeoffs, between existing and prospective model configurations. For example, when a manufacturer considers replacing an underlying foundation model, the current and proposed versions of the complete device can be evaluated against common clinical safety criteria, including relevant failure modes and population-specific requirements. Because Circuit Breaker Labs produces numerical, explainable results, manufacturers can identify potential improvements and regressions before implementing the change.
The same approach can help detect unexpected changes introduced by third-party model providers. By periodically evaluating the deployed device against an established baseline, manufacturers can identify behavioral regressions even when they have not received notification of an underlying modification.
We believe recurring evaluation should be a reasonable expectation for GenAI-enabled medical devices, with frequency proportionate to clinical risk and the pace of system change. Monthly or quarterly assessments may be appropriate for some applications, while high-consequence systems or those dependent on frequently updated foundation models may warrant more frequent or event-triggered testing.
Importantly, periodic testing should complement, rather than replace, evaluations conducted before known model changes are deployed. Manufacturers should also consider technical and contractual mechanisms such as model-version pinning, advance update notifications, access to previous versions, and rollback capabilities where available.
Advances in automated, clinically-informed evaluation make broad regression testing substantially more scalable than exclusively manual approaches. This creates an opportunity to maintain rigorous oversight without imposing the same cost and operational burden as repeated full-scale clinical studies.
Ultimately, the appropriate regulatory expectation should not be that manufacturers can prevent every third-party model change. It should be that they maintain reliable mechanisms to detect clinically significant changes, assess their consequences, and take timely corrective action, regardless of whether the change originated within or outside their organization.
Other Topics
Question 25: Foundation Model Documentation
Are voluntary Foundation Model Master Files (MAFs) sufficient to support regulatory review, or should the FDA consider alternative evidence mechanisms?
We appreciate the potential value of voluntary Foundation MAFs, particularly as a mechanism for improving transparency about underlying models used in medical devices. However, we are concerned that voluntary disclosure alone may not provide sufficiently current or application-specific evidence of safety.
Foundation model developers may be slow to continually update regulatory documentation, particularly when they are not themselves developing the downstream medical device and may not be attuned to the specific risks.
A more fundamental limitation is that foundation model safety is not equivalent to device safety. A foundation model's performance depends substantially on how it is incorporated into a specific application. System prompts, retrieval-augmented generation, conversational memory, and guardrails can all introduce new vulnerabilities or alter existing ones.
Similarly, the clinical safety requirements for a behavioral health application may differ substantially from those for a medication management system, even when both use the same underlying foundation model.
For these reasons, we would not recommend relying on a voluntary MAF as the primary mechanism for establishing or maintaining the safety of downstream devices.
Instead, CDRH could consider a complementary framework involving ongoing independent foundation model evaluations and application-specific testing.
For example, qualified independent evaluation organizations could periodically compare major foundation models against common healthcare-relevant competencies, including clinical knowledge, safety-critical recognition, uncertainty communication, and consistency of behavior.
Such comparative evaluations, “bakeoffs” as Circuit Breaker Labs refers to them, could provide useful evidence about the relative strengths, limitations, and changing performance of different underlying models.
CDRH could consider maintaining or referencing a body of independently-generated evaluation evidence, with clear documentation of model versions, assessment methods, intended contexts, and known limitations. This could provide reviewers with additional information when assessing devices built on commonly used foundation models.
At Circuit Breaker Labs, comparative evaluation of different model configurations is one approach we use to help characterize clinically relevant performance differences. We believe a similar methodology could contribute to a broader evidence ecosystem without requiring foundation model developers to disclose proprietary training information or pending non-public updates.
However, these comparisons should not be interpreted as general authorization or certification of an underlying model for medical use. A model that performs well on common healthcare benchmarks may still behave unsafely when integrated into a particular device or altered via RAG, etc.
If this approach is taken, we recommend treating foundation model evidence as a useful reference that can inform, but not replace, application-specific evaluation.
Voluntary MAFs could still contribute important information, including version identification, known limitations, major safety-relevant changes, and update-notification practices. These disclosures would be most useful when combined with proactive, recurring behavioral testing rather than relied upon as a static representation of model safety.
The objective should be to establish a regulatory evidence framework that remains informative even when model providers make unanticipated changes, use cases evolve, or the performance characteristics of a previously evaluated model shift over time.
Question 26: Agentic AI
What additional safety requirements and oversight are necessary for autonomous medical AI systems that perform multi-step actions and use external tools?
Agentic generative AI-enabled devices warrant additional scrutiny because their risks may emerge across sequences of decisions and actions rather than from any individual output.
This is particularly important for systems capable of using external tools, maintaining conversational context, modifying records, initiating clinical workflows, or performing actions with limited opportunities for human review.
We recommend that CDRH place particular emphasis on high-volume, clinically-informed, trajectory-level evaluation in both premarket and postmarket settings.
Generative AI systems can produce variable outputs in response to similar inputs, and agentic systems introduce further complexity because small differences in an early decision may propagate through subsequent actions. Evaluating a small number of isolated examples is therefore unlikely to adequately characterize the range of potential failures.
Manufacturers should evaluate realistic multi-step scenarios, including variations in the order of information, incomplete patient histories, conflicting inputs, changing clinical conditions, and extended interactions in which important safety information may become less salient.
A central concern is context degradation. An agent may initially recognize a clinically significant contraindication but subsequently fail to account for it when performing an action many steps later. Likewise, an agent may correctly execute individual tool calls while producing an unsafe overall outcome because it has lost track of the clinical context or failed to reassess whether its planned actions remain appropriate.
These risks are not hypothetical properties of tool use alone. They arise from the interaction between the underlying generative system, its context-management mechanisms, the tools it can access, and the clinical consequences of its decisions.
Our experience conducting trajectory-based evaluations in behavioral health has reinforced the importance of assessing complete interactions. The order in which information is introduced, the progression of the user's circumstances, and the accumulated conversational context can materially affect how safety safeguards are invoked and whether a system maintains appropriate boundaries.
We recommend that acceptance criteria therefore address not only the correctness of individual responses or tool calls, but also the safety of the complete action sequence.
Relevant competencies should include persistent recognition of clinically significant information, adherence to authorized scope, appropriate human-approval checkpoints, resistance to prompt injection, handling of erroneous tool outputs, and the ability to stop or safely recover when circumstances change.
Equally important is the scientific reliability of the evaluation process itself. Agentic systems produce complex, open-ended trajectories, but this does not mean their performance must be assessed through subjective or irreproducible judgments.
Evaluation methods can incorporate clinical context and nuanced behavioral criteria while producing structured, quantitative, and auditable results. We recommend that manufacturers demonstrate how their scoring methods distinguish clinically significant failures, how reliably those methods agree with qualified clinical judgment, and whether repeated evaluation of identical trajectories produces consistent results.
A zero-variance evaluator should assign the same score to an identical trajectory. Such reproducibility is particularly valuable for identifying safety regressions across device versions and postmarket updates. Generative LLM-as-a-judge approaches may provide useful qualitative assessments, but their variability, systematic biases, and potential correlated errors should be characterized before they are relied upon for safety-critical determinations. A changing evaluation method can make it difficult to determine whether a device has actually improved or deteriorated.
Finally, we recommend that agentic device evaluation combine large-scale synthetic testing with clinically representative confirmation and continuing postmarket reassessment. Testing should be sufficiently broad to explore complex and uncommon trajectories, while statistical claims should account for the representativeness and dependence of generated scenarios.
The objective should be to establish reproducible evidence that clinically significant safeguards persist across complex action sequences, changing contexts, and evolving system configurations.
As generative AI becomes increasingly capable of acting rather than merely providing information, the importance of measuring these properties will only grow.
FDA Comment
Considerations for the Regulation of Generative AI-Enabled Medical Devices
Circuit Breaker Labs | FDA Public Comment • October 19, 2026
Assessing Risk
Question 1: Risk Framework
Does the proposed two-axis risk framework adequately capture the risks of generative AI medical devices, or should additional factors be considered?
The proposed two-axis framework provides a useful foundation, particularly because it recognizes risk as a continuum. However, device activity and the consequences of an incorrect output do not fully characterize the risks of generative AI-enabled medical devices. We recommend retaining the two axes while introducing additional axes.
First, risk depends on the characteristics of the intended user. A patient, a practicing clinician, a specialist, and a clinician in training may interact with the same system differently and have different abilities to recognize an incorrect response. Even within these populations, age, health literacy, clinical complexity, and patterns of use can materially affect risk. For instance, adolescent users may use language that differs substantially from the system's expected inputs. Risk assessment should consider foreseeable use behaviors rather than assuming all users interact with the device as intended.
Second, informational functions warrant greater differentiation. A resource navigator that directs users to established clinical information differs meaningfully from a system that synthesizes medication information in the context of an individual's treatment history. Even without an explicit recommendation, personalized information may influence clinical decisions. The extent to which a system incorporates patient-specific information should therefore modify its risk classification.
Third, the presence of human supervision should not automatically be treated as equivalent to effective oversight. A clinician who reviews every output before it reaches a patient provides a different safeguard from one who retrospectively reviews a small sample of interactions. Risk assessment should consider the timing, frequency, coverage, clinical competence, and intervention authority associated with human oversight. Retrospective sampling may detect systematic problems but cannot prevent harm from an individual output already delivered.
Fourth, traceability and robustness should be considered together. A system may correctly handle a clinically significant concept when expressed using formal terminology but respond differently when the same information is expressed colloquially. For example, requests involving “acetaminophen” and “Tylenol” should not produce materially different safety-critical behavior merely because of the terminology used, but we have observed materially different behavior in our evaluations at Circuit Breaker Labs. Traceability to appropriate clinical sources is valuable, but it does not itself establish that the system will apply those sources consistently.
Rather than adding numerous independent axes, CDRH could establish a small set of risk modifiers and associated testing expectations. This would preserve the flexibility of the proposed framework while giving manufacturers more predictable, application-specific requirements.
Question 2: Informational Functions
How should the FDA assess the risks of AI-generated information based on its wording, personalization, context, and influence on user decisions?
The degree to which an informational function directs behavior cannot be assessed solely through explicit instructions or recommendations. Risk also depends on how the information is communicated, the user's circumstances, and the degree of authority or interpersonal trust the system may implicitly convey.
A particularly important consideration is anthropomorphism, the attribution of human characteristics or understanding to an AI system. A system that speaks in a highly personal, emotionally affirming, or companion-like manner may encourage users to place greater confidence in its outputs, even when those outputs are nominally non-directive. Generative systems may also mirror or reinforce a user's beliefs and emotions, a behavior that can be inappropriate in certain clinical contexts.
We recommend a principle of minimum necessary anthropomorphism: manufacturers should justify personifying design features according to their intended clinical benefit and assess whether those features introduce undue trust, emotional dependence, or inappropriate behavioral influence. This need not prohibit first-person language or empathetic communication. Indeed, some therapeutic applications may benefit from carefully designed relational interactions. The important distinction is between communication that supports appropriate engagement and communication that encourages users to overestimate the system's understanding, competence, or authority.
CDRH could provide greater predictability through a behavioral assessment framework considering the use of patient-specific information, implied authority, emotional influence, and persistence of advice across repeated interactions.
We also recommend that manufacturers evaluate these properties across relevant patient populations, clinical conditions, and conversational contexts. Attachment-related risks may emerge gradually, particularly during extended interactions, and may not be apparent from isolated responses.
Rather than requiring manufacturers to eliminate all personalization, CDRH should encourage empirical evaluation of how personalization affects comprehension, trust calibration, and downstream decision-making.
Question 3: Patient-Facing AI
How should the FDA account for the unique risks of delivering AI-generated clinical information directly to patients rather than clinicians?
Patient-facing generative AI applications offer substantial potential benefits, including more timely access to health information, improved patient understanding, and support when a clinician is not immediately available. These benefits should not be discounted merely because patients lack formal clinical training.
However, patient-facing systems may require stronger safeguards because the recipient may not be able to identify an incorrect output or recognize when an apparently routine question involves clinically significant risk.
We recommend an approach in which multiple complementary safeguards address different failure modes. These should include clinically-grounded constraints, appropriate uncertainty communication, reliable escalation pathways, and testing against foreseeable real-world inputs.
An important consideration is the reliability of the safeguards themselves. Some systems use generative language models to evaluate or moderate the outputs of other generative models (‘LLM-as-a-judge’). Such approaches can introduce additional model-dependent error or variability. For safety-critical judgments, manufacturers should characterize the evaluator's sensitivity, specificity, reproducibility, variance, and susceptibility to adversarial or out-of-distribution inputs.
Patient-facing systems should also be evaluated using realistic input variation, including misspellings, colloquialisms, incomplete medical histories, contradictory information, differences in health literacy, and population-specific language. The goal should be reliable performance across the intended population.
Clinical claims should be supported by appropriate, current evidence, with uncertainty clearly communicated where evidence is limited. In clinically ambiguous or high-consequence circumstances, systems should be able to defer or escalate appropriately without over-refusing routine informational requests.
Importantly, these safeguards should be assessed through observable behavior, not solely through documentation that a safety policy or guardrail has been implemented.
Question 4: Clinical Specialization
How should differences between generalist and specialist physicians affect the risk assessment and safeguards of clinician-facing AI?
The distinction between generalist and specialist clinicians is important, but risk assessment should reflect the actual clinical workflow. We recommend involving both generalist and specialist clinicians in the development and evaluation of relevant GenAI-enabled functions. Generalists can help assess how a system is likely to be used in routine practice, while specialists can identify subtle errors, contraindications, and clinically significant failure modes that might otherwise go undetected.
This is particularly important for systems that provide specialized information to users without corresponding specialist training. A physician may appropriately rely on an AI system for support outside their primary field but may be less equipped to recognize when its recommendations are subtly incorrect.
Behavioral health provides a useful example. An AI system that repeatedly reassures or validates a user's concerns may appear an empathetic sounding board. However, in the context of obsessive-compulsive disorder, repetitive reassurance can reinforce maladaptive cycles. Similarly, uncritical affirmation of delusional beliefs may be clinically inappropriate. Evaluations developed without relevant specialist input may fail to distinguish appropriate emotional support from harmful reinforcement.
We recommend that clinical specialists contribute not only to reviewing outputs, but also to defining failure categories, evaluation criteria, and clinically appropriate safeguards. Generalist clinicians should participate in realistic workflow testing to determine whether the system communicates these distinctions effectively to its actual users.
For applications whose intended use population could reasonably span multiple clinical domains, the evaluation should also include relevant adjacent specialties and common comorbid conditions. A device intended for one specialty may create risks outside that specialty through medication interactions, overlapping symptoms, or treatment decisions.
Question 5: Multi-Turn Conversations
How should the FDA evaluate risks that emerge or change over the course of extended AI conversations?
Multi-turn conversational devices introduce risks that cannot be adequately characterized through isolated prompts. Safety-critical behavior may change over the course of a conversation as information accumulates, earlier details become less salient, and the user's presentation evolves.
One particularly important concern is context pollution. Clinically consequential information introduced early, e.g., during patient history, may no longer influence later outputs appropriately because of context-window limitations, summarization errors, retrieval failures, or the model's inconsistent use of available information.
For example, a patient may disclose a serious medication allergy early in an interaction, then return to medication-related questions much later. The continued absence of an allergic reaction does not make that contraindication irrelevant. Similarly, a system instructed to provide non-directive information may gradually transition toward individualized treatment recommendations as conversation context accumulates.
We recommend that CDRH encourage evaluation of complete conversational trajectories, including scenarios in which risks emerge gradually, relevant information is disclosed indirectly, users resist escalation, and important details are separated by substantial conversational distance.
Testing should examine whether safety-critical constraints persist despite changes in topic, user framing, and conversation length. Manufacturers should characterize the reliability of their memory mechanisms, context-management strategies, and safety controls under realistic and adversarial conditions.
Where appropriate, safety-critical information should be preserved in structured, persistent representations rather than relying exclusively on the generative model to recall and correctly prioritize it. Such representations should themselves be tested for accuracy, appropriate updates, and protection against conflicting or misleading information.
We further recommend distinguishing turn-level safety from trajectory-level safety. A system may produce individually plausible responses while its cumulative behavior becomes unsafe. The unit of evaluation must therefore include the conversation and its clinical consequences, not merely each response in isolation.
Question 6: Care Escalation
How should manufacturers measure and balance the risks of escalating care too frequently versus failing to escalate when necessary?
Under-escalation and over-escalation should be characterized separately because their consequences differ across clinical contexts and cannot always be reduced to a common error measure.
Repeated unnecessary escalation may also discourage patients from engaging with the system or seeking appropriate care in the future.
We recommend defining escalation thresholds according to the severity of potential harm, time sensitivity, uncertainty, availability of clinical supervision, and reversibility of an incorrect decision.
For instance, a clinician-supervised informational function may appropriately provide differentiated risk information without reflexively escalating every ambiguous presentation. Conversely, an autonomous function operating without timely human review may require a lower threshold for escalation in high-consequence situations.
Clinical prevalence and patient heterogeneity should also inform evaluation. In rare or highly variable conditions, the absence of a familiar symptom pattern may provide limited reassurance. However, rarity alone should not automatically mandate escalation. The relevant consideration is whether uncertainty and potential harm justify additional review.
CDRH should encourage manufacturers to report under-escalation and over-escalation as distinct outcomes, with clinically justified thresholds for each. Evaluations should include true emergencies, borderline presentations, and clearly non-urgent cases.
Particular attention should be given to escalation timing. A system that ultimately recommends emergency care may still fail if it first provides several rounds of inappropriate reassurance or delays the recommendation beyond a clinically meaningful point.
Competency-Based Premarket Evaluation
Question 7: Competency-Based Evaluation
Is a framework combining device benchmarking with clinical confirmation appropriate for evaluating generative AI medical devices?
We support the proposed competency-based approach, particularly the distinction between scalable non-clinical benchmarking and clinical confirmation under representative conditions.
That said, representative, high-volume testing of open-ended generative systems is not impractical. As a point of evidence: Circuit Breaker Labs has designed an evaluation system for monitoring of clinical AI that can run hundreds of thousands of dynamic, multi-turn tests each day, building meaningful clinical evidence at rapid speed.
Importantly, competency should not be understood as a static property established at the time of initial evaluation. Generative AI-enabled devices may change because of modifications to underlying models, prompts, guardrails, retrieval systems, user interfaces, or other components. Even changes not explicitly intended to affect clinical or safety performance may alter behavior in a clinically significant way.
We therefore recommend treating competency-based evaluation as the foundation of a continuous assurance process. Premarket benchmarking would establish a characterized baseline, clinical confirmation would provide evidence of performance in representative use, and postmarket reassessment would determine whether that competency is maintained as the system evolves.
The core competencies and acceptance criteria remain stable to support meaningful longitudinal comparisons. At the same time, individual test cases should be refreshed and expanded to account for emerging failure modes, changing clinical practices, and previously unidentified populations or conditions.
This would preserve standardization without allowing initial benchmark performance to become a substitute for continued evidence of safety and effectiveness.
Question 8: Risk-Based Evidence
How should a device's risk classification determine the amount and type of evidence required before FDA approval?
The two-axis framework could provide the foundation for determining the rigor and scope of premarket evidence, but the additional risk modifiers discussed in Question 1 should also inform what manufacturers are expected to demonstrate.
In particular, CDRH should consider the intended user, degree of personalization, clinical supervision model, patient population, potential for extended interactions, and clinical dependencies extending beyond the device's primary specialty.
A device providing general cardiovascular education, for example, should not necessarily face the same evidentiary expectations as one generating individualized cardiovascular treatment recommendations. Likewise, a cardiology-focused device may need to demonstrate appropriate handling of pulmonary disease, renal impairment, and other relevant comorbidities when they influence its intended function.
We recommend that manufacturers establish a risk-to-evidence mapping before testing. This would identify which clinically significant behaviors must be evaluated, which populations and deployment conditions must be represented, and what acceptance thresholds are justified for each category.
This mapping would also provide a reproducible starting point for postmarket evaluation. When device components change, manufacturers could determine which risk categories may have been affected and conduct targeted reassessment, while retaining broader regression testing for unexpected effects.
A standardized approach to mapping risk categories to evidence requirements could improve regulatory predictability without imposing identical testing obligations on devices with materially different risk profiles.
Question 9: Benchmarking Standards
Does the proposed benchmarking framework adequately assess clinical knowledge, reasoning, safety, communication, and generalizability, or are additional standards needed?
The proposed benchmarking elements provide a strong foundation. We particularly support the inclusion of safety-critical escalation, scope adherence, uncertainty communication, robustness, subgroup performance, and agentic capabilities.
However, we recommend some refinements.
First, CDRH should distinguish reusable evaluation criteria from reusable test instances. Standardized clinical competencies, scoring definitions, and acceptance criteria are essential for comparability. Repeated reliance on a fixed, publicly available set of prompts or data is less reliable, particularly when models or applications are often optimized against those exact examples. The evaluation framework should therefore support standardized evaluation metrics applied to both reference cases and independently generated, sequestered, or refreshed scenarios.
Second, clinically significant context retention should be an explicit competency. Devices should demonstrate that they maintain and appropriately apply relevant patient information across extended interactions, including contraindications, allergies, comorbidities, and changes in clinical status. This competency spans clinical reasoning, safety, robustness, and agentic behavior, but its importance warrants explicit consideration.
Third, anthropomorphism, inappropriate agreeableness, and automation bias should receive greater attention. A system may provide factually accurate information while encouraging excessive reliance or reinforcing harmful beliefs. These risks can emerge through communication style, emotional affirmation, and repeated interactions, particularly in behavioral health applications.
Fourth, clinical provenance should be assessed. Where a device presents information as clinically established, users should be able to distinguish established guidance from emerging evidence, inference, or uncertain claims. Unsupported assertions of authority should be considered separately from ordinary factual errors.
Finally, the proposed elements overlap in clinically meaningful ways. For example, an agent that reorders a medication without adequately considering a relevant contraindication may demonstrate failures of clinical proficiency, context handling, and agentic execution simultaneously.
We do not recommend treating these categories as mutually exclusive. Instead, CDRH could retain the organizing structure while allowing individual failures to map to multiple competencies, with careful safeguards against double-counting in aggregate scores.
Question 10: Benchmark Validity
How can manufacturers demonstrate that benchmark performance reliably predicts real-world safety without overfitting or introducing conflicts of interest?
CDRH identifies a central challenge in generative AI evaluation: performance on a known benchmark does not necessarily predict performance under novel, real-world conditions.
We recommend distinguishing the clinical construct being measured from the particular examples used to measure it. Clinical risk categories, evaluation criteria, and acceptance thresholds should be clearly specified and sufficiently stable to support reproducibility. However, the scenarios used to assess those should be capable of evolving as new failure modes and patterns of use emerge.
A robust approach could combine common reference benchmarks, independent sequestered cases, and dynamically generated scenarios tailored to a device's intended use. This would allow manufacturers to assess established competencies while reducing overreliance on a finite set of recognizable inputs.
The construct validity of such benchmarks should be supported through qualified clinical review, explicit links to clinically meaningful failure mechanisms, testing across realistic input variation, and comparison with independent clinical adjudication where appropriate.
We also recommend evaluating measurement reliability separately from model performance. When a scoring method produces inconsistent judgments across equivalent inputs or repeated assessments, an apparent change in device performance may reflect measurement variation rather than an actual change in safety. Manufacturers should characterize the repeatability, agreement with qualified adjudicators, and limitations of their evaluation methods.
Independent evaluation organizations can help identify blind spots arising from sponsor-developed test suites, particularly when they collaborate with clinical specialists. However, independence should not preclude device-specific customization. A useful structure could include a common set of clinically defined risks supplemented by independently developed scenarios addressing the manufacturer's intended use, population, and known vulnerabilities.
Manufacturers should document what risks their systems may encounter, why those risks matter, and independent evaluators can assess handling of these risks and provide detail on how the scenarios represent intended use, and what important uncertainties remain.
Finally, benchmark maintenance should be treated as a continuing scientific activity. As underlying models and clinical applications evolve, evaluation methods must be updated without sacrificing reproducibility or longitudinal comparability.
Question 11: Clinical Confirmation
How should manufacturers select appropriate clinical validation methods based on device risk, intended use, and real-world patient interactions?
We support a risk-proportionate approach to clinical confirmation that allows manufacturers to select methods appropriate to their device's intended use, degree of autonomy, and potential consequences of failure. However, we recommend that the selection process begin with a structured characterization of clinical risks rather than a determination of which study design is most convenient.
Manufacturers could first describe their device's risk profile using the proposed two-axis framework, supplemented by the additional considerations discussed in our earlier responses. This assessment should identify both the situations the device is intended to address and foreseeable circumstances in which its outputs could cause harm.
For example, a cardiology-focused system may not explicitly address pulmonary disease, yet many patients presenting with cardiovascular concerns also have respiratory conditions or overlapping symptoms. A risk assessment that excludes these clinical intersections could overlook consequential failure modes. CDRH could request additional evidence in areas where the intended function creates foreseeable dependencies on adjacent clinical specialties.
We recommend that manufacturers document these anticipated risks and provide corresponding performance evidence, including findings from qualified independent evaluators where appropriate. Independent evaluators can help identify blind spots that may be overlooked when manufacturers develop their own evaluation criteria, while clinical specialists and generalist physicians can provide complementary perspectives on clinical complexity and actual patterns of use.
The anticipated distribution of real-world inputs should inform both test coverage and the amount of testing performed. A device expected to support large volumes of patient interactions should be evaluated across a sufficiently broad range of scenarios to characterize uncommon but clinically significant failures.
High-throughput automated testing makes it feasible to examine many more scenarios than traditional manual evaluation alone.
For perspective, our work at Circuit Breaker Labs has focused on scalable, clinically-informed testing methods capable of evaluating large numbers of interactions efficiently. Such methods can supplement clinical confirmation by systematically examining variation in patient characteristics, language, comorbidities, and conversational trajectories.
It is also important to distinguish reproducibility of the evaluation method from variability in the device being evaluated. Reproducible scoring methods can make longitudinal comparisons more reliable, but their clinical validity must still be established through appropriate expert agreement and other evidence.
Finally, we recommend a staged approach in which scalable benchmarking identifies areas of uncertainty that then inform more resource-intensive confirmation methods. For instance, failures involving complex comorbidities discovered through synthetic testing could guide targeted retrospective analysis, standardized patient interactions, or shadow deployment.
This would allow manufacturers to use clinical evidence more efficiently while directing greater scrutiny toward the conditions most likely to produce significant harm.
Question 12: Statistical Reliability
How should manufacturers establish statistically meaningful performance estimates when combining synthetic testing, real-world data, and clinical evidence?
Statistically meaningful performance measurement is particularly challenging for generative AI-enabled devices because outputs may vary across repeated interactions, clinically significant failures may be uncommon, and the real-world distribution of patient inputs may be incompletely characterized.
We recommend that CDRH encourage large-scale, risk-stratified evaluation supported by clinically-informed synthetic data, while clearly distinguishing performance under synthetic testing conditions from estimated real-world performance.
Real-world patient data is essential for understanding how individuals actually interact with medical AI, including the conditions they present with, their language patterns, and the frequency of different clinical scenarios. However, particularly for emerging conversational and patient-facing applications, available data may not adequately represent the full range of potential interactions.
Privacy-preserving analysis of real-world data can help characterize these patterns and inform the construction of synthetic test scenarios. Clinical experts can then assess whether the generated cases are medically plausible, reflect meaningful patterns of presentation, and cover relevant populations and conditions.
Importantly, synthetic data should not merely reproduce the observed distribution of existing patient interactions. Real-world datasets may systematically underrepresent rare conditions, underserved populations, atypical symptoms, or clinically important scenarios that have not yet been encountered. Synthetic testing creates an opportunity to intentionally examine such situations before they result in real-world harm.
Advances in automated evaluation have made testing at the scale of hundreds of thousands or millions of interactions technically feasible, as evidenced by the in-market tool developed by Circuit Breaker Labs. We believe this capability should be leveraged to characterize performance across a broad range of clinically meaningful scenarios, including variations in patient language, age, comorbidities, and conversational histories.
We recommend separating three components of performance evidence:
We also recommend reporting uncertainty intervals, clinically relevant subgroup performance, and failure severity rather than relying exclusively on a single aggregate score.
The objective should be to make statistically rigorous evaluation more accessible and scalable.
Question 13: Synthetic Data
Where is synthetic data most useful for evaluating medical AI, and how can manufacturers prevent it from reproducing existing model biases and blind spots?
We believe synthetic data can make substantial contributions across clinical domains, device functions, and patient populations, particularly where real-world data is insufficient to characterize the full range of foreseeable inputs.
Conversational and behavioral health applications illustrate both the need for synthetic data and the importance of developing it carefully. Unlike many structured clinical datasets, patient-facing conversations can include indirect disclosures, evolving symptoms, incomplete histories, informal language, and substantial variation in how individuals describe similar experiences. Two patients presenting with the same underlying clinical concern may use entirely different language, and relatively small wording differences may influence whether an AI system recognizes a safety-critical condition or responds appropriately.
For example, a patient describing significant psychological distress may not use formal diagnostic terminology or explicitly identify the condition they are experiencing. A system evaluated primarily on textbook descriptions may perform well on a benchmark while failing to recognize clinically meaningful variations encountered in practice.
Synthetic testing offers a way to systematically explore these differences. It can generate variations in wording, age-related language, clinical context, comorbidities, symptom progression, and conversation length while preserving a clinically specified underlying scenario.
However, the clinical risks being tested should not be done solely by the same class of generative models being evaluated (‘LLM-as-a-judge approaches’). Otherwise judging may reproduce overlapping assumptions or blind spots.
Clinical experts should establish the relevant risk categories and behavioral expectations. Synthetic scenarios can then be generated through complementary techniques, including structured simulation and steering. Finally, outputs should demonstrate reliability and clinical validity.
At Circuit Breaker Labs, our approach emphasizes clinically-informed risk criteria and reproducible evaluation methods rather than solely on one generative model to independently produce scenarios and another to judge the results. We believe this separation is an important design principle for developing trustworthy evaluation infrastructure.
The involvement of clinicians experienced with underrepresented populations is especially important. Available patient datasets may not fully reflect how individuals across age groups, dialects, health literacy levels, cultural contexts, and co-occurring conditions describe their concerns. Specialists and generalists can identify clinically meaningful variations that would otherwise be absent from test-generation processes.
We also recommend that synthetic datasets be continuously expanded and validated as new evidence emerges from clinical practice, postmarket monitoring, patient feedback, and independent evaluations.
The appropriate goal is therefore not to replace real-world evidence, but to extend the range of conditions under which safety can be examined before patients encounter those conditions themselves.
Question 14: Performance Comparators
What clinical standards, physician expertise, or human-AI performance benchmarks should determine whether an AI device performs acceptably?
We recommend that performance comparators and acceptance criteria be determined according to a device's intended use, level of autonomy, patient population, and clinical consequences of failure.
The proposed axis framework could provide a starting point for establishing baseline expectations. Manufacturers should define the functions within the device's intended scope, the behaviors that would constitute unacceptable deviations from that scope, and the clinical competencies required for safe operation.
Minimum requirements could include recognition of clinically significant contraindications, appropriate escalation in emergencies, adherence to the device's intended scope, and avoidance of clearly harmful recommendations. These should be grounded in applicable clinical standards and established with input from qualified specialists.
Comparative performance could then be assessed against clinicians or human-AI teams appropriate to the intended workflow.
The relative contributions of generalist and specialist physicians deserve particular consideration.
We therefore recommend involving both groups in evaluation design while assigning responsibility for defining particular clinical acceptance criteria to professionals with qualifications appropriate to the clinical question.
The degree of human involvement in analogous real-world operation should also determine which comparator is most relevant. If a system is intended to support a clinician who reviews and acts on its recommendations, the combined performance of the clinician and AI may be a central endpoint. If the system makes decisions or takes actions autonomously, its independent performance may be assessed.
Patient-facing systems present an additional consideration. The absence of professional review at the point of interaction may necessitate more conservative safety requirements, stronger escalation pathways, and ongoing monitoring. However, retrospective human review should not automatically be treated as equivalent to clinician supervision before a consequential response reaches a patient.
Finally, because generative systems can produce many different clinically acceptable responses, scoring should focus on whether the response satisfies clinically meaningful criteria rather than whether it matches a single reference answer.
Question 15: Real-World Alternatives
Should AI medical devices be evaluated against the care patients would realistically receive without them, including delayed care or no treatment?
We strongly support consideration of the care or intervention that would realistically occur in the absence of a generative AI-enabled device. This is particularly important for applications intended to improve access-to-care.
In behavioral health, for example, the relevant alternative to an AI-enabled support tool may not always be an appointment with a licensed therapist. For some individuals, the alternative may be delayed care, limited access to clinical information, reliance on informal support, or no intervention at all.
Evaluating a device exclusively against an idealized healthcare scenario may therefore underestimate its potential benefits, particularly for populations facing substantial barriers to access.
However, we recommend distinguishing benefit relative to the available alternative from the device's absolute safety. A device may offer meaningful benefits where no timely care would otherwise be available while still creating risks that require independent assessment.
One particularly important question is whether an AI interaction introduces a clinically significant harm that would not otherwise have occurred.
For example, in behavioral health, a user experiencing ordinary sadness may seek conversational support. An AI system that amplifies emotional distress, or encourages inappropriate dependence could potentially worsen the user's condition, even if it never provides an explicitly dangerous instruction.
Conversely, a well-designed system may provide helpful information, encourage appropriate care-seeking, or offer beneficial support during periods when professional care is unavailable.
We recommend that clinical confirmation consider whether the device produces a net benefit, neutral effect, or additional harm relative to a realistic counterfactual, while separately measuring serious adverse outcomes that should not be offset by average benefit.
Potential comparators could include usual care, delayed specialist review, unaided clinical judgment, available digital resources, or no intervention, depending on the intended use and actual deployment setting.
For devices performing clinical reasoning or decision-making, comparison with qualified clinicians remains important. Specialist judgment may be particularly relevant for assessing complex cases, while generalist judgment can help characterize the decisions likely to occur earlier in a patient's care pathway.
When feasible, studies could examine not only whether the device's decision agrees with clinical experts, but also whether exposure to the device changes patient understanding, decision-making, care-seeking behavior, or clinical outcomes.
Such assessments should account for baseline patient characteristics and other factors that may influence outcomes independently of the AI system.
We believe this approach would allow FDA to recognize the substantial potential of generative AI to expand access to healthcare without overlooking the possibility that some applications create new harms rather than simply performing existing clinical functions less effectively.
Question 16: Independent Evaluation
What role should qualified independent third parties play in AI medical device evaluation, and how can their participation preserve competition and innovation?
We strongly support the involvement of qualified independent third parties throughout the competency-based evaluation process. Blind spots are natural, and independent scrutiny is particularly valuable for identifying failure modes that may not be apparent to the organizations developing a device.
Independent evaluators can contribute during premarket development by identifying vulnerabilities, assessing clinically significant behaviors, and providing evidence that informs design decisions and safeguards. This need not be limited to a final assessment immediately before submission. Early and iterative evaluation can help manufacturers address risks while products are still being developed, rather than discovering them after deployment.
The resulting evidence can also establish a baseline for postmarket monitoring. By evaluating devices against consistent clinical factors over time, independent organizations can help manufacturers identify emerging vulnerabilities, quantify regressions, and determine whether modifications have affected previously demonstrated competencies.
We recommend that qualifications reflect both clinical and technical expertise. Evaluators should demonstrate the ability to define clinically meaningful failure categories, incorporate perspectives from relevant specialists and patient populations, conduct sufficiently broad and high-volume testing, and produce reproducible, auditable findings. Technical testing capacity alone should not be sufficient; evaluation methods must also demonstrate clinical validity.
Importantly, independent evaluation need not require access to a manufacturer's proprietary models, training data, system prompts, or internal architecture. For example, at Circuit Breaker Labs, we conduct external evaluations through API endpoint-based access to deployed systems, allowing manufacturers to retain control of their underlying intellectual property. This approach can reduce barriers to participation while enabling independent assessment of observable device behavior.
Finally, we suggest that CDRH avoid creating an evaluation framework that inadvertently concentrates market access among a small number of organizations. Qualified for-profit companies, nonprofit laboratories, academic groups, and other independent evaluators should be able to participate under transparent, technology-neutral qualification criteria.
Independence should be defined by the integrity of the evaluation process, not the legal or commercial structure of the organization conducting it. Such an approach would encourage competition in evaluation methodologies, encourage innovation in the evaluation process itself, protect manufacturer intellectual property, and support the development of safety infrastructure capable of keeping pace with innovation.
Question 17: Multimodal AI
How should competency-based evaluation adapt to medical AI systems that process voice, images, and other modalities beyond text?
We believe the competency-based approach can be applied across different model architectures, provided that the competencies are defined according to the device's intended clinical function rather than the particular architecture of its underlying model.
The central regulatory question should remain whether the final deployed device performs its intended function safely and effectively under foreseeable conditions. However, multimodal devices introduce additional failure modes that may not be adequately captured through evaluations designed primarily for text-based systems.
For example, a voice-enabled device may respond appropriately to an accurate transcript while failing to recognize clinically significant information when it is spoken with an unfamiliar accent, interrupted, or affected by background noise. We have observed this with our voice red-teaming offering at Circuit Breaker Labs. Similarly, an image-based system may accurately interpret a clear clinical image but behave differently when the image is incomplete, poorly illuminated, or inconsistent with accompanying patient information.
We therefore recommend evaluating both individual components and the complete end-to-end clinical interaction. Component-level testing can help identify whether a failure originated in speech recognition, image interpretation, retrieval, reasoning, or another process. However, successful performance by individual components does not necessarily establish that the complete device will behave safely when those components interact.
An especially important competency is the handling of conflicting information across modalities. For example, a patient's verbal description of a condition may conflict with an image, a physiological measurement, or a previously documented clinical history. Evaluation should establish whether the device appropriately recognizes uncertainty, gathers additional information, or defers when necessary rather than arbitrarily prioritizing one input source.
Multimodal evaluation should also consider patient-specific communication patterns, including accents, dialects, age-related language differences, interrupted speech, and changes in how users present information across the course of an interaction.
Multi-participant red-teaming, an offering Circuit Breaker Labs, is also critical to consider as some AI-enabled medical devices may involve one or more patients and clinicians interfacing with an AI product, where risk is between human-AI and human-human.
For generative models, additional considerations may include the accuracy of modeled state transitions, the system's response to uncertainty, and the consequences of incorrect predictions over multiple steps. A plausible simulated future state does not necessarily establish that the predicted outcome is clinically reliable.
These systems should be assessed through realistic scenarios in which initial information may be incomplete, assumptions may be incorrect, and errors can accumulate over time.
Postmarket Monitoring
Question 18: Premarket vs. Postmarket Evidence
Under what circumstances should the FDA accept greater uncertainty before approval in exchange for stronger postmarket monitoring?
We recognize the rationale for considering greater reliance on postmarket monitoring, particularly because generative AI-enabled medical devices may operate in settings where real-world usage patterns and the full distribution of clinical risks are not yet well characterized.
However, postmarket monitoring should complement a meaningful premarket evidence baseline. This is particularly important for emerging patient-facing applications. In behavioral health, for example, available real-world data may not adequately represent the range of ways patients discuss sensitive conditions. Privacy constraints, limited data availability, and the evolving nature of human-AI interactions can make it difficult to construct comprehensive evaluation datasets from historical conversations alone.
We therefore recommend that manufacturers use clinically-informed synthetic scenarios to supplement available real-world evidence during premarket evaluation. These scenarios can systematically explore clinically plausible variations in patient characteristics, language, symptom presentation, comorbidities, and conversational trajectories.
High-volume testing is especially valuable in this context because it enables manufacturers to examine a broad range of foreseeable failures before users encounter them. Clinical experts should be involved in defining the risks being tested and evaluating whether synthetic scenarios reflect meaningful clinical situations.
Postmarket monitoring can then assess whether performance remains consistent with that baseline, whether changes introduce new vulnerabilities, and whether previously unidentified failure modes emerge as real-world evidence accumulates.
Greater reliance on postmarket evidence may be reasonable when the device has demonstrated adequate initial safety, the remaining uncertainties can be monitored meaningfully, and mechanisms exist to identify and remediate clinically significant problems in a timely manner.
For patient-facing conversational systems, human psychology introduces further considerations. A device's outputs may influence trust, emotional reliance, care-seeking behavior, and subsequent disclosures. These effects may develop across extended interactions and should not be assumed to be captured by conventional single-turn benchmarks.
We therefore recommend an approach in which robust premarket synthetic and clinical evaluation establishes a safety baseline, while continuous monitoring expands the evidence base and detects deviations after deployment.
This would support innovation without treating patients as the primary means of discovering risks that could reasonably have been identified before market entry. Humans should not be Guinea pigs.
Question 19: Postmarket Monitoring
How frequently should medical AI devices be reassessed, and what changes or events should trigger additional safety testing?
We support periodic re-benchmarking, sample-based clinician review, and performance degradation monitoring. However, we recommend that CDRH explicitly distinguish event-triggered reassessment from time-based reassessment, and consider both essential components of postmarket oversight.
Event-triggered evaluations are valuable when manufacturers make known changes that may affect device performance. Examples include modifications to the underlying foundation model, prompts, retrieval architecture, safety controls, guardrails, or clinical functionality.
However, manufacturers may not always know when their systems have changed in clinically significant ways. Third-party model providers may modify hosted services, previously unrecognized vulnerabilities may emerge, and changes in deployment conditions may affect behavior without a clearly identifiable software modification.
For this reason, postmarket monitoring should not depend exclusively on manufacturers recognizing that a change has occurred.
We recommend recurring re-benchmarking against established clinical safety criteria, even when no known modification has taken place. Such evaluations could identify clinically meaningful regressions, newly emerging vulnerabilities, or changes in the consistency of safety-critical behavior.
The appropriate cadence should be proportionate to the device's risk profile, anticipated usage, clinical setting, and rate of technical change. Monthly or quarterly evaluation may be appropriate for some applications, while higher-risk functions may require more frequent monitoring.
Clinical expertise should remain central throughout this process. Clinicians should contribute to defining evaluation criteria, identifying emerging clinical risks, and adjudicating cases where automated methods cannot reliably determine the clinical significance of an output.
We further recommend that monitoring include refreshed or independently developed challenge cases to help detect vulnerabilities that may not appear in previously used benchmarks.
The objective should be to create a monitoring system that identifies both regressions in known competencies and failures that were not anticipated during initial evaluation.
Question 20: Automated AI Oversight
How should the FDA evaluate the accuracy, reliability, and limitations of automated systems used to monitor other medical AI devices?
Machine-based supervisory systems can absolutely play an important role in scalable postmarket monitoring, particularly where devices produce large volumes of open-ended outputs and tests are necessary for generative AI systems, but cannot feasibly be reviewed individually by clinicians.
However, we caution against assuming that one generative model can reliably validate another. LLM-as-a-judge approaches exhibit several documented limitations, including:
These concerns are especially significant when supervisory systems are used to determine whether a safety-critical behavior has improved, degraded, or remained unchanged.
A broader concern is correlated error. When the device under evaluation and the supervisory model share similar assumptions, training influences, or failure modes, agreement between the two may create a false impression of independent validation.
We therefore recommend that CDRH encourage the use of complementary and independently validated evaluation mechanisms, including non-generative methods, specialized scoring systems, and qualified clinician review where appropriate.
Non-generative evaluation techniques can be designed to produce reproducible numerical scores while assessing clinically relevant contextual features. Circuit Breaker Labs, for example, develops clinically-informed machine-based evaluation methods that do not rely on a general-purpose LLM to assign final judgments.
We recommend that manufacturers characterize the reliability of any supervisory evaluation system by requesting repeated assessments of identical outputs.
CDRH should also consider evaluation throughput and operational scalability. A central advantage of machine-based supervision is the ability to evaluate substantially larger volumes of interactions than manual review alone. However, this advantage is meaningful only when the evaluation method is sufficiently accurate, reliable, and clinically appropriate.
Ultimately, the supervisory system should itself be treated as a measurement instrument whose validity, reliability, and limitations must be demonstrated. A changing device cannot be meaningfully monitored using an evaluation method whose own behavior is insufficiently characterized.
Question 21: Stakeholder Accountability
How should clinicians, healthcare institutions, independent evaluators, and other stakeholders contribute to postmarket monitoring while keeping manufacturers accountable?
All these stakeholders have valuable roles in postmarket monitoring, but CDRH should also recognize the important contributions of independent, for-profit evaluation organizations. Given the pace at which generative AI is advancing, a sustainable safety ecosystem must be capable of innovating at a comparable pace to for-profit manufacturers. Commercial evaluation organizations can have incentives to invest in scalable testing infrastructure, develop new methodologies, and rapidly adapt to emerging risks and new capabilities. Their involvement should be encouraged alongside academic institutions, professional societies, clinicians, and other stakeholders.
Shared participation in monitoring need not mean shared accountability for remediation. Manufacturers can remain responsible for defining their device's risk profile and demonstrating that appropriate safeguards are in place. Independent evaluators and other stakeholders can supplement this work by identifying previously unrecognized vulnerabilities, assessing performance against established clinical criteria, and detecting emerging risks through postmarket monitoring and reporting those back to manufacturers.
When monitoring identifies a clinically significant issue, the manufacturer should be responsible for investigating and addressing it within a timeframe proportionate to its severity. Where remediation cannot be completed within that timeframe, manufacturers should document a corrective action plan, interim risk mitigations, and an expected resolution timeline, with reporting to FDA where applicable. Particularly serious risks may require immediate intervention rather than awaiting completion of a remediation plan.
This approach allows different stakeholders to contribute their expertise without obscuring responsibility. Independent organizations can identify and characterize failures, clinicians and professional societies can inform their clinical significance, and manufacturers remain accountable for maintaining the safety and effectiveness of their devices.
Question 22: Re-Benchmarking After Changes
How should the extent of safety testing and regulatory review depend on the type and potential consequences of modifications to an AI device?
We recommend that post-deployment reassessment be proportional to the potential clinical consequences of a modification rather than solely its technical complexity or the manufacturer's intended purpose for the change.
A relatively minor software modification may have outsized implications for device safety. For example, changes to a system prompt, knowledge retrieval mechanism, context management strategy, or underlying foundation model may alter how a device interprets contraindications, recognizes clinical emergencies, or responds to patients outside its intended scope.
We propose a risk-based re-benchmarking framework built on the premarket competency assessment.
Manufacturers should maintain a mapping between their device's components, intended functions, clinically-significant failure categories, and established acceptance criteria. When a modification occurs, this mapping can help identify the competencies potentially affected and the evidence needed to demonstrate continued performance.
For example, a change to the user interface that does not materially affect clinical information or user decision-making may warrant limited reassessment. A change to the underlying language model, safety architecture, or conversational memory system may warrant broader testing, including multi-turn scenarios and relevant clinical subpopulations.
However, even targeted reassessment should be supplemented by cadenced safety-critical regression tests. Generative systems can exhibit unexpected interactions between components, and manufacturers should not assume that a modification affects only the functions they intended to change.
We recommend distinguishing regression testing from discovery-oriented testing. Regression testing determines whether established competencies have been maintained. Discovery-oriented testing uses refreshed, independently developed, or adversarial scenarios to identify new vulnerabilities that may emerge unexpectedly.
Changes that fall within a prospectively defined and appropriately validated modification framework may be suitable for management through a PCCP, consistent with applicable regulatory requirements. Changes that introduce new intended functions, substantially alter risk exposure, or exceed the boundaries of an established PCCP may require additional evidence or FDA review.
Importantly, the amount of evidence should be informed by the severity and detectability of potential harm, the degree of clinical autonomy, and the availability of effective downstream safeguards.
This approach would make re-benchmarking more efficient and predictable while recognizing that the safety impact of a modification cannot always be inferred from the size or apparent significance of the underlying engineering change.
Question 23: Unpredictable Modifications
How should predetermined change control plans (PCCPs) account for AI system changes that cannot be fully anticipated in advance?
PCCPs are valuable for modifications manufacturers can reasonably anticipate, including changes to underlying models, prompts, retrieval systems, or other components that may affect device performance. However, generative AI-enabled devices present an additional challenge: clinically meaningful behavioral changes may occur even when the manufacturer has not knowingly implemented a modification.
For example, a third-party foundation model provider may update a hosted model without providing sufficient visibility into the change. Even when the model identifier remains unchanged, differences in backend configuration or other dependencies may potentially affect performance.
We therefore recommend complementing modification-triggered PCCPs with proactive, recurring evaluation. Rather than testing only when a manufacturer anticipates that something has changed, a device could be periodically reassessed against its established clinical competencies and safety requirements, regardless of whether a modification has been identified.
This would create two complementary safeguards: predefined change-control procedures for anticipated modifications, and ongoing monitoring to detect unexpected changes, previously unidentified vulnerabilities, or performance degradation.
Importantly, the monitoring framework should define which safety-critical behaviors must remain consistent, how changes will be measured, what constitutes a clinically significant regression, and what corrective actions would follow. Evaluation methods should produce numerical, sufficiently reproducible results that can be tracked to distinguish changes in device performance from variability in the measurement process itself.
This approach would allow manufacturers to prepare for the reality that not every future change can be predicted, without requiring them to prespecify every possible modification. The central principle is that manufacturers may not know what will change, but they can define what safety-critical behaviors must continue to hold.
Question 24: Third-Party Model Updates
How can medical device manufacturers detect and address safety risks introduced by updates to third-party foundation models?
Third-party foundation model dependencies create a significant challenge because manufacturers may not control, anticipate, or even be notified of changes that affect their device's clinical performance.
We recommend that manufacturers implement ongoing behavioral monitoring and comparative evaluations of their deployed systems. Independent evaluation organizations, including companies such as Circuit Breaker Labs, can support this process through recurring safety assessments and controlled comparisons, or bakeoffs, between existing and prospective model configurations. For example, when a manufacturer considers replacing an underlying foundation model, the current and proposed versions of the complete device can be evaluated against common clinical safety criteria, including relevant failure modes and population-specific requirements. Because Circuit Breaker Labs produces numerical, explainable results, manufacturers can identify potential improvements and regressions before implementing the change.
The same approach can help detect unexpected changes introduced by third-party model providers. By periodically evaluating the deployed device against an established baseline, manufacturers can identify behavioral regressions even when they have not received notification of an underlying modification.
We believe recurring evaluation should be a reasonable expectation for GenAI-enabled medical devices, with frequency proportionate to clinical risk and the pace of system change. Monthly or quarterly assessments may be appropriate for some applications, while high-consequence systems or those dependent on frequently updated foundation models may warrant more frequent or event-triggered testing.
Importantly, periodic testing should complement, rather than replace, evaluations conducted before known model changes are deployed. Manufacturers should also consider technical and contractual mechanisms such as model-version pinning, advance update notifications, access to previous versions, and rollback capabilities where available.
Advances in automated, clinically-informed evaluation make broad regression testing substantially more scalable than exclusively manual approaches. This creates an opportunity to maintain rigorous oversight without imposing the same cost and operational burden as repeated full-scale clinical studies.
Ultimately, the appropriate regulatory expectation should not be that manufacturers can prevent every third-party model change. It should be that they maintain reliable mechanisms to detect clinically significant changes, assess their consequences, and take timely corrective action, regardless of whether the change originated within or outside their organization.
Other Topics
Question 25: Foundation Model Documentation
Are voluntary Foundation Model Master Files (MAFs) sufficient to support regulatory review, or should the FDA consider alternative evidence mechanisms?
We appreciate the potential value of voluntary Foundation MAFs, particularly as a mechanism for improving transparency about underlying models used in medical devices. However, we are concerned that voluntary disclosure alone may not provide sufficiently current or application-specific evidence of safety.
Foundation model developers may be slow to continually update regulatory documentation, particularly when they are not themselves developing the downstream medical device and may not be attuned to the specific risks.
A more fundamental limitation is that foundation model safety is not equivalent to device safety. A foundation model's performance depends substantially on how it is incorporated into a specific application. System prompts, retrieval-augmented generation, conversational memory, and guardrails can all introduce new vulnerabilities or alter existing ones.
Similarly, the clinical safety requirements for a behavioral health application may differ substantially from those for a medication management system, even when both use the same underlying foundation model.
For these reasons, we would not recommend relying on a voluntary MAF as the primary mechanism for establishing or maintaining the safety of downstream devices.
Instead, CDRH could consider a complementary framework involving ongoing independent foundation model evaluations and application-specific testing.
For example, qualified independent evaluation organizations could periodically compare major foundation models against common healthcare-relevant competencies, including clinical knowledge, safety-critical recognition, uncertainty communication, and consistency of behavior.
Such comparative evaluations, “bakeoffs” as Circuit Breaker Labs refers to them, could provide useful evidence about the relative strengths, limitations, and changing performance of different underlying models.
CDRH could consider maintaining or referencing a body of independently-generated evaluation evidence, with clear documentation of model versions, assessment methods, intended contexts, and known limitations. This could provide reviewers with additional information when assessing devices built on commonly used foundation models.
At Circuit Breaker Labs, comparative evaluation of different model configurations is one approach we use to help characterize clinically relevant performance differences. We believe a similar methodology could contribute to a broader evidence ecosystem without requiring foundation model developers to disclose proprietary training information or pending non-public updates.
However, these comparisons should not be interpreted as general authorization or certification of an underlying model for medical use. A model that performs well on common healthcare benchmarks may still behave unsafely when integrated into a particular device or altered via RAG, etc.
If this approach is taken, we recommend treating foundation model evidence as a useful reference that can inform, but not replace, application-specific evaluation.
Voluntary MAFs could still contribute important information, including version identification, known limitations, major safety-relevant changes, and update-notification practices. These disclosures would be most useful when combined with proactive, recurring behavioral testing rather than relied upon as a static representation of model safety.
The objective should be to establish a regulatory evidence framework that remains informative even when model providers make unanticipated changes, use cases evolve, or the performance characteristics of a previously evaluated model shift over time.
Question 26: Agentic AI
What additional safety requirements and oversight are necessary for autonomous medical AI systems that perform multi-step actions and use external tools?
Agentic generative AI-enabled devices warrant additional scrutiny because their risks may emerge across sequences of decisions and actions rather than from any individual output.
This is particularly important for systems capable of using external tools, maintaining conversational context, modifying records, initiating clinical workflows, or performing actions with limited opportunities for human review.
We recommend that CDRH place particular emphasis on high-volume, clinically-informed, trajectory-level evaluation in both premarket and postmarket settings.
Generative AI systems can produce variable outputs in response to similar inputs, and agentic systems introduce further complexity because small differences in an early decision may propagate through subsequent actions. Evaluating a small number of isolated examples is therefore unlikely to adequately characterize the range of potential failures.
Manufacturers should evaluate realistic multi-step scenarios, including variations in the order of information, incomplete patient histories, conflicting inputs, changing clinical conditions, and extended interactions in which important safety information may become less salient.
A central concern is context degradation. An agent may initially recognize a clinically significant contraindication but subsequently fail to account for it when performing an action many steps later. Likewise, an agent may correctly execute individual tool calls while producing an unsafe overall outcome because it has lost track of the clinical context or failed to reassess whether its planned actions remain appropriate.
These risks are not hypothetical properties of tool use alone. They arise from the interaction between the underlying generative system, its context-management mechanisms, the tools it can access, and the clinical consequences of its decisions.
Our experience conducting trajectory-based evaluations in behavioral health has reinforced the importance of assessing complete interactions. The order in which information is introduced, the progression of the user's circumstances, and the accumulated conversational context can materially affect how safety safeguards are invoked and whether a system maintains appropriate boundaries.
We recommend that acceptance criteria therefore address not only the correctness of individual responses or tool calls, but also the safety of the complete action sequence.
Relevant competencies should include persistent recognition of clinically significant information, adherence to authorized scope, appropriate human-approval checkpoints, resistance to prompt injection, handling of erroneous tool outputs, and the ability to stop or safely recover when circumstances change.
Equally important is the scientific reliability of the evaluation process itself. Agentic systems produce complex, open-ended trajectories, but this does not mean their performance must be assessed through subjective or irreproducible judgments.
Evaluation methods can incorporate clinical context and nuanced behavioral criteria while producing structured, quantitative, and auditable results. We recommend that manufacturers demonstrate how their scoring methods distinguish clinically significant failures, how reliably those methods agree with qualified clinical judgment, and whether repeated evaluation of identical trajectories produces consistent results.
A zero-variance evaluator should assign the same score to an identical trajectory. Such reproducibility is particularly valuable for identifying safety regressions across device versions and postmarket updates. Generative LLM-as-a-judge approaches may provide useful qualitative assessments, but their variability, systematic biases, and potential correlated errors should be characterized before they are relied upon for safety-critical determinations. A changing evaluation method can make it difficult to determine whether a device has actually improved or deteriorated.
Finally, we recommend that agentic device evaluation combine large-scale synthetic testing with clinically representative confirmation and continuing postmarket reassessment. Testing should be sufficiently broad to explore complex and uncommon trajectories, while statistical claims should account for the representativeness and dependence of generated scenarios.
The objective should be to establish reproducible evidence that clinically significant safeguards persist across complex action sequences, changing contexts, and evolving system configurations.
As generative AI becomes increasingly capable of acting rather than merely providing information, the importance of measuring these properties will only grow.
FDA Comment
Considerations for the Regulation of Generative AI-Enabled Medical Devices
Circuit Breaker Labs | FDA Public Comment • October 19, 2026
Assessing Risk
Question 1: Risk Framework
Does the proposed two-axis risk framework adequately capture the risks of generative AI medical devices, or should additional factors be considered?
The proposed two-axis framework provides a useful foundation, particularly because it recognizes risk as a continuum. However, device activity and the consequences of an incorrect output do not fully characterize the risks of generative AI-enabled medical devices. We recommend retaining the two axes while introducing additional axes.
First, risk depends on the characteristics of the intended user. A patient, a practicing clinician, a specialist, and a clinician in training may interact with the same system differently and have different abilities to recognize an incorrect response. Even within these populations, age, health literacy, clinical complexity, and patterns of use can materially affect risk. For instance, adolescent users may use language that differs substantially from the system's expected inputs. Risk assessment should consider foreseeable use behaviors rather than assuming all users interact with the device as intended.
Second, informational functions warrant greater differentiation. A resource navigator that directs users to established clinical information differs meaningfully from a system that synthesizes medication information in the context of an individual's treatment history. Even without an explicit recommendation, personalized information may influence clinical decisions. The extent to which a system incorporates patient-specific information should therefore modify its risk classification.
Third, the presence of human supervision should not automatically be treated as equivalent to effective oversight. A clinician who reviews every output before it reaches a patient provides a different safeguard from one who retrospectively reviews a small sample of interactions. Risk assessment should consider the timing, frequency, coverage, clinical competence, and intervention authority associated with human oversight. Retrospective sampling may detect systematic problems but cannot prevent harm from an individual output already delivered.
Fourth, traceability and robustness should be considered together. A system may correctly handle a clinically significant concept when expressed using formal terminology but respond differently when the same information is expressed colloquially. For example, requests involving “acetaminophen” and “Tylenol” should not produce materially different safety-critical behavior merely because of the terminology used, but we have observed materially different behavior in our evaluations at Circuit Breaker Labs. Traceability to appropriate clinical sources is valuable, but it does not itself establish that the system will apply those sources consistently.
Rather than adding numerous independent axes, CDRH could establish a small set of risk modifiers and associated testing expectations. This would preserve the flexibility of the proposed framework while giving manufacturers more predictable, application-specific requirements.
Question 2: Informational Functions
How should the FDA assess the risks of AI-generated information based on its wording, personalization, context, and influence on user decisions?
The degree to which an informational function directs behavior cannot be assessed solely through explicit instructions or recommendations. Risk also depends on how the information is communicated, the user's circumstances, and the degree of authority or interpersonal trust the system may implicitly convey.
A particularly important consideration is anthropomorphism, the attribution of human characteristics or understanding to an AI system. A system that speaks in a highly personal, emotionally affirming, or companion-like manner may encourage users to place greater confidence in its outputs, even when those outputs are nominally non-directive. Generative systems may also mirror or reinforce a user's beliefs and emotions, a behavior that can be inappropriate in certain clinical contexts.
We recommend a principle of minimum necessary anthropomorphism: manufacturers should justify personifying design features according to their intended clinical benefit and assess whether those features introduce undue trust, emotional dependence, or inappropriate behavioral influence. This need not prohibit first-person language or empathetic communication. Indeed, some therapeutic applications may benefit from carefully designed relational interactions. The important distinction is between communication that supports appropriate engagement and communication that encourages users to overestimate the system's understanding, competence, or authority.
CDRH could provide greater predictability through a behavioral assessment framework considering the use of patient-specific information, implied authority, emotional influence, and persistence of advice across repeated interactions.
We also recommend that manufacturers evaluate these properties across relevant patient populations, clinical conditions, and conversational contexts. Attachment-related risks may emerge gradually, particularly during extended interactions, and may not be apparent from isolated responses.
Rather than requiring manufacturers to eliminate all personalization, CDRH should encourage empirical evaluation of how personalization affects comprehension, trust calibration, and downstream decision-making.
Question 3: Patient-Facing AI
How should the FDA account for the unique risks of delivering AI-generated clinical information directly to patients rather than clinicians?
Patient-facing generative AI applications offer substantial potential benefits, including more timely access to health information, improved patient understanding, and support when a clinician is not immediately available. These benefits should not be discounted merely because patients lack formal clinical training.
However, patient-facing systems may require stronger safeguards because the recipient may not be able to identify an incorrect output or recognize when an apparently routine question involves clinically significant risk.
We recommend an approach in which multiple complementary safeguards address different failure modes. These should include clinically-grounded constraints, appropriate uncertainty communication, reliable escalation pathways, and testing against foreseeable real-world inputs.
An important consideration is the reliability of the safeguards themselves. Some systems use generative language models to evaluate or moderate the outputs of other generative models (‘LLM-as-a-judge’). Such approaches can introduce additional model-dependent error or variability. For safety-critical judgments, manufacturers should characterize the evaluator's sensitivity, specificity, reproducibility, variance, and susceptibility to adversarial or out-of-distribution inputs.
Patient-facing systems should also be evaluated using realistic input variation, including misspellings, colloquialisms, incomplete medical histories, contradictory information, differences in health literacy, and population-specific language. The goal should be reliable performance across the intended population.
Clinical claims should be supported by appropriate, current evidence, with uncertainty clearly communicated where evidence is limited. In clinically ambiguous or high-consequence circumstances, systems should be able to defer or escalate appropriately without over-refusing routine informational requests.
Importantly, these safeguards should be assessed through observable behavior, not solely through documentation that a safety policy or guardrail has been implemented.
Question 4: Clinical Specialization
How should differences between generalist and specialist physicians affect the risk assessment and safeguards of clinician-facing AI?
The distinction between generalist and specialist clinicians is important, but risk assessment should reflect the actual clinical workflow. We recommend involving both generalist and specialist clinicians in the development and evaluation of relevant GenAI-enabled functions. Generalists can help assess how a system is likely to be used in routine practice, while specialists can identify subtle errors, contraindications, and clinically significant failure modes that might otherwise go undetected.
This is particularly important for systems that provide specialized information to users without corresponding specialist training. A physician may appropriately rely on an AI system for support outside their primary field but may be less equipped to recognize when its recommendations are subtly incorrect.
Behavioral health provides a useful example. An AI system that repeatedly reassures or validates a user's concerns may appear an empathetic sounding board. However, in the context of obsessive-compulsive disorder, repetitive reassurance can reinforce maladaptive cycles. Similarly, uncritical affirmation of delusional beliefs may be clinically inappropriate. Evaluations developed without relevant specialist input may fail to distinguish appropriate emotional support from harmful reinforcement.
We recommend that clinical specialists contribute not only to reviewing outputs, but also to defining failure categories, evaluation criteria, and clinically appropriate safeguards. Generalist clinicians should participate in realistic workflow testing to determine whether the system communicates these distinctions effectively to its actual users.
For applications whose intended use population could reasonably span multiple clinical domains, the evaluation should also include relevant adjacent specialties and common comorbid conditions. A device intended for one specialty may create risks outside that specialty through medication interactions, overlapping symptoms, or treatment decisions.
Question 5: Multi-Turn Conversations
How should the FDA evaluate risks that emerge or change over the course of extended AI conversations?
Multi-turn conversational devices introduce risks that cannot be adequately characterized through isolated prompts. Safety-critical behavior may change over the course of a conversation as information accumulates, earlier details become less salient, and the user's presentation evolves.
One particularly important concern is context pollution. Clinically consequential information introduced early, e.g., during patient history, may no longer influence later outputs appropriately because of context-window limitations, summarization errors, retrieval failures, or the model's inconsistent use of available information.
For example, a patient may disclose a serious medication allergy early in an interaction, then return to medication-related questions much later. The continued absence of an allergic reaction does not make that contraindication irrelevant. Similarly, a system instructed to provide non-directive information may gradually transition toward individualized treatment recommendations as conversation context accumulates.
We recommend that CDRH encourage evaluation of complete conversational trajectories, including scenarios in which risks emerge gradually, relevant information is disclosed indirectly, users resist escalation, and important details are separated by substantial conversational distance.
Testing should examine whether safety-critical constraints persist despite changes in topic, user framing, and conversation length. Manufacturers should characterize the reliability of their memory mechanisms, context-management strategies, and safety controls under realistic and adversarial conditions.
Where appropriate, safety-critical information should be preserved in structured, persistent representations rather than relying exclusively on the generative model to recall and correctly prioritize it. Such representations should themselves be tested for accuracy, appropriate updates, and protection against conflicting or misleading information.
We further recommend distinguishing turn-level safety from trajectory-level safety. A system may produce individually plausible responses while its cumulative behavior becomes unsafe. The unit of evaluation must therefore include the conversation and its clinical consequences, not merely each response in isolation.
Question 6: Care Escalation
How should manufacturers measure and balance the risks of escalating care too frequently versus failing to escalate when necessary?
Under-escalation and over-escalation should be characterized separately because their consequences differ across clinical contexts and cannot always be reduced to a common error measure.
Repeated unnecessary escalation may also discourage patients from engaging with the system or seeking appropriate care in the future.
We recommend defining escalation thresholds according to the severity of potential harm, time sensitivity, uncertainty, availability of clinical supervision, and reversibility of an incorrect decision.
For instance, a clinician-supervised informational function may appropriately provide differentiated risk information without reflexively escalating every ambiguous presentation. Conversely, an autonomous function operating without timely human review may require a lower threshold for escalation in high-consequence situations.
Clinical prevalence and patient heterogeneity should also inform evaluation. In rare or highly variable conditions, the absence of a familiar symptom pattern may provide limited reassurance. However, rarity alone should not automatically mandate escalation. The relevant consideration is whether uncertainty and potential harm justify additional review.
CDRH should encourage manufacturers to report under-escalation and over-escalation as distinct outcomes, with clinically justified thresholds for each. Evaluations should include true emergencies, borderline presentations, and clearly non-urgent cases.
Particular attention should be given to escalation timing. A system that ultimately recommends emergency care may still fail if it first provides several rounds of inappropriate reassurance or delays the recommendation beyond a clinically meaningful point.
Competency-Based Premarket Evaluation
Question 7: Competency-Based Evaluation
Is a framework combining device benchmarking with clinical confirmation appropriate for evaluating generative AI medical devices?
We support the proposed competency-based approach, particularly the distinction between scalable non-clinical benchmarking and clinical confirmation under representative conditions.
That said, representative, high-volume testing of open-ended generative systems is not impractical. As a point of evidence: Circuit Breaker Labs has designed an evaluation system for monitoring of clinical AI that can run hundreds of thousands of dynamic, multi-turn tests each day, building meaningful clinical evidence at rapid speed.
Importantly, competency should not be understood as a static property established at the time of initial evaluation. Generative AI-enabled devices may change because of modifications to underlying models, prompts, guardrails, retrieval systems, user interfaces, or other components. Even changes not explicitly intended to affect clinical or safety performance may alter behavior in a clinically significant way.
We therefore recommend treating competency-based evaluation as the foundation of a continuous assurance process. Premarket benchmarking would establish a characterized baseline, clinical confirmation would provide evidence of performance in representative use, and postmarket reassessment would determine whether that competency is maintained as the system evolves.
The core competencies and acceptance criteria remain stable to support meaningful longitudinal comparisons. At the same time, individual test cases should be refreshed and expanded to account for emerging failure modes, changing clinical practices, and previously unidentified populations or conditions.
This would preserve standardization without allowing initial benchmark performance to become a substitute for continued evidence of safety and effectiveness.
Question 8: Risk-Based Evidence
How should a device's risk classification determine the amount and type of evidence required before FDA approval?
The two-axis framework could provide the foundation for determining the rigor and scope of premarket evidence, but the additional risk modifiers discussed in Question 1 should also inform what manufacturers are expected to demonstrate.
In particular, CDRH should consider the intended user, degree of personalization, clinical supervision model, patient population, potential for extended interactions, and clinical dependencies extending beyond the device's primary specialty.
A device providing general cardiovascular education, for example, should not necessarily face the same evidentiary expectations as one generating individualized cardiovascular treatment recommendations. Likewise, a cardiology-focused device may need to demonstrate appropriate handling of pulmonary disease, renal impairment, and other relevant comorbidities when they influence its intended function.
We recommend that manufacturers establish a risk-to-evidence mapping before testing. This would identify which clinically significant behaviors must be evaluated, which populations and deployment conditions must be represented, and what acceptance thresholds are justified for each category.
This mapping would also provide a reproducible starting point for postmarket evaluation. When device components change, manufacturers could determine which risk categories may have been affected and conduct targeted reassessment, while retaining broader regression testing for unexpected effects.
A standardized approach to mapping risk categories to evidence requirements could improve regulatory predictability without imposing identical testing obligations on devices with materially different risk profiles.
Question 9: Benchmarking Standards
Does the proposed benchmarking framework adequately assess clinical knowledge, reasoning, safety, communication, and generalizability, or are additional standards needed?
The proposed benchmarking elements provide a strong foundation. We particularly support the inclusion of safety-critical escalation, scope adherence, uncertainty communication, robustness, subgroup performance, and agentic capabilities.
However, we recommend some refinements.
First, CDRH should distinguish reusable evaluation criteria from reusable test instances. Standardized clinical competencies, scoring definitions, and acceptance criteria are essential for comparability. Repeated reliance on a fixed, publicly available set of prompts or data is less reliable, particularly when models or applications are often optimized against those exact examples. The evaluation framework should therefore support standardized evaluation metrics applied to both reference cases and independently generated, sequestered, or refreshed scenarios.
Second, clinically significant context retention should be an explicit competency. Devices should demonstrate that they maintain and appropriately apply relevant patient information across extended interactions, including contraindications, allergies, comorbidities, and changes in clinical status. This competency spans clinical reasoning, safety, robustness, and agentic behavior, but its importance warrants explicit consideration.
Third, anthropomorphism, inappropriate agreeableness, and automation bias should receive greater attention. A system may provide factually accurate information while encouraging excessive reliance or reinforcing harmful beliefs. These risks can emerge through communication style, emotional affirmation, and repeated interactions, particularly in behavioral health applications.
Fourth, clinical provenance should be assessed. Where a device presents information as clinically established, users should be able to distinguish established guidance from emerging evidence, inference, or uncertain claims. Unsupported assertions of authority should be considered separately from ordinary factual errors.
Finally, the proposed elements overlap in clinically meaningful ways. For example, an agent that reorders a medication without adequately considering a relevant contraindication may demonstrate failures of clinical proficiency, context handling, and agentic execution simultaneously.
We do not recommend treating these categories as mutually exclusive. Instead, CDRH could retain the organizing structure while allowing individual failures to map to multiple competencies, with careful safeguards against double-counting in aggregate scores.
Question 10: Benchmark Validity
How can manufacturers demonstrate that benchmark performance reliably predicts real-world safety without overfitting or introducing conflicts of interest?
CDRH identifies a central challenge in generative AI evaluation: performance on a known benchmark does not necessarily predict performance under novel, real-world conditions.
We recommend distinguishing the clinical construct being measured from the particular examples used to measure it. Clinical risk categories, evaluation criteria, and acceptance thresholds should be clearly specified and sufficiently stable to support reproducibility. However, the scenarios used to assess those should be capable of evolving as new failure modes and patterns of use emerge.
A robust approach could combine common reference benchmarks, independent sequestered cases, and dynamically generated scenarios tailored to a device's intended use. This would allow manufacturers to assess established competencies while reducing overreliance on a finite set of recognizable inputs.
The construct validity of such benchmarks should be supported through qualified clinical review, explicit links to clinically meaningful failure mechanisms, testing across realistic input variation, and comparison with independent clinical adjudication where appropriate.
We also recommend evaluating measurement reliability separately from model performance. When a scoring method produces inconsistent judgments across equivalent inputs or repeated assessments, an apparent change in device performance may reflect measurement variation rather than an actual change in safety. Manufacturers should characterize the repeatability, agreement with qualified adjudicators, and limitations of their evaluation methods.
Independent evaluation organizations can help identify blind spots arising from sponsor-developed test suites, particularly when they collaborate with clinical specialists. However, independence should not preclude device-specific customization. A useful structure could include a common set of clinically defined risks supplemented by independently developed scenarios addressing the manufacturer's intended use, population, and known vulnerabilities.
Manufacturers should document what risks their systems may encounter, why those risks matter, and independent evaluators can assess handling of these risks and provide detail on how the scenarios represent intended use, and what important uncertainties remain.
Finally, benchmark maintenance should be treated as a continuing scientific activity. As underlying models and clinical applications evolve, evaluation methods must be updated without sacrificing reproducibility or longitudinal comparability.
Question 11: Clinical Confirmation
How should manufacturers select appropriate clinical validation methods based on device risk, intended use, and real-world patient interactions?
We support a risk-proportionate approach to clinical confirmation that allows manufacturers to select methods appropriate to their device's intended use, degree of autonomy, and potential consequences of failure. However, we recommend that the selection process begin with a structured characterization of clinical risks rather than a determination of which study design is most convenient.
Manufacturers could first describe their device's risk profile using the proposed two-axis framework, supplemented by the additional considerations discussed in our earlier responses. This assessment should identify both the situations the device is intended to address and foreseeable circumstances in which its outputs could cause harm.
For example, a cardiology-focused system may not explicitly address pulmonary disease, yet many patients presenting with cardiovascular concerns also have respiratory conditions or overlapping symptoms. A risk assessment that excludes these clinical intersections could overlook consequential failure modes. CDRH could request additional evidence in areas where the intended function creates foreseeable dependencies on adjacent clinical specialties.
We recommend that manufacturers document these anticipated risks and provide corresponding performance evidence, including findings from qualified independent evaluators where appropriate. Independent evaluators can help identify blind spots that may be overlooked when manufacturers develop their own evaluation criteria, while clinical specialists and generalist physicians can provide complementary perspectives on clinical complexity and actual patterns of use.
The anticipated distribution of real-world inputs should inform both test coverage and the amount of testing performed. A device expected to support large volumes of patient interactions should be evaluated across a sufficiently broad range of scenarios to characterize uncommon but clinically significant failures.
High-throughput automated testing makes it feasible to examine many more scenarios than traditional manual evaluation alone.
For perspective, our work at Circuit Breaker Labs has focused on scalable, clinically-informed testing methods capable of evaluating large numbers of interactions efficiently. Such methods can supplement clinical confirmation by systematically examining variation in patient characteristics, language, comorbidities, and conversational trajectories.
It is also important to distinguish reproducibility of the evaluation method from variability in the device being evaluated. Reproducible scoring methods can make longitudinal comparisons more reliable, but their clinical validity must still be established through appropriate expert agreement and other evidence.
Finally, we recommend a staged approach in which scalable benchmarking identifies areas of uncertainty that then inform more resource-intensive confirmation methods. For instance, failures involving complex comorbidities discovered through synthetic testing could guide targeted retrospective analysis, standardized patient interactions, or shadow deployment.
This would allow manufacturers to use clinical evidence more efficiently while directing greater scrutiny toward the conditions most likely to produce significant harm.
Question 12: Statistical Reliability
How should manufacturers establish statistically meaningful performance estimates when combining synthetic testing, real-world data, and clinical evidence?
Statistically meaningful performance measurement is particularly challenging for generative AI-enabled devices because outputs may vary across repeated interactions, clinically significant failures may be uncommon, and the real-world distribution of patient inputs may be incompletely characterized.
We recommend that CDRH encourage large-scale, risk-stratified evaluation supported by clinically-informed synthetic data, while clearly distinguishing performance under synthetic testing conditions from estimated real-world performance.
Real-world patient data is essential for understanding how individuals actually interact with medical AI, including the conditions they present with, their language patterns, and the frequency of different clinical scenarios. However, particularly for emerging conversational and patient-facing applications, available data may not adequately represent the full range of potential interactions.
Privacy-preserving analysis of real-world data can help characterize these patterns and inform the construction of synthetic test scenarios. Clinical experts can then assess whether the generated cases are medically plausible, reflect meaningful patterns of presentation, and cover relevant populations and conditions.
Importantly, synthetic data should not merely reproduce the observed distribution of existing patient interactions. Real-world datasets may systematically underrepresent rare conditions, underserved populations, atypical symptoms, or clinically important scenarios that have not yet been encountered. Synthetic testing creates an opportunity to intentionally examine such situations before they result in real-world harm.
Advances in automated evaluation have made testing at the scale of hundreds of thousands or millions of interactions technically feasible, as evidenced by the in-market tool developed by Circuit Breaker Labs. We believe this capability should be leveraged to characterize performance across a broad range of clinically meaningful scenarios, including variations in patient language, age, comorbidities, and conversational histories.
We recommend separating three components of performance evidence:
We also recommend reporting uncertainty intervals, clinically relevant subgroup performance, and failure severity rather than relying exclusively on a single aggregate score.
The objective should be to make statistically rigorous evaluation more accessible and scalable.
Question 13: Synthetic Data
Where is synthetic data most useful for evaluating medical AI, and how can manufacturers prevent it from reproducing existing model biases and blind spots?
We believe synthetic data can make substantial contributions across clinical domains, device functions, and patient populations, particularly where real-world data is insufficient to characterize the full range of foreseeable inputs.
Conversational and behavioral health applications illustrate both the need for synthetic data and the importance of developing it carefully. Unlike many structured clinical datasets, patient-facing conversations can include indirect disclosures, evolving symptoms, incomplete histories, informal language, and substantial variation in how individuals describe similar experiences. Two patients presenting with the same underlying clinical concern may use entirely different language, and relatively small wording differences may influence whether an AI system recognizes a safety-critical condition or responds appropriately.
For example, a patient describing significant psychological distress may not use formal diagnostic terminology or explicitly identify the condition they are experiencing. A system evaluated primarily on textbook descriptions may perform well on a benchmark while failing to recognize clinically meaningful variations encountered in practice.
Synthetic testing offers a way to systematically explore these differences. It can generate variations in wording, age-related language, clinical context, comorbidities, symptom progression, and conversation length while preserving a clinically specified underlying scenario.
However, the clinical risks being tested should not be done solely by the same class of generative models being evaluated (‘LLM-as-a-judge approaches’). Otherwise judging may reproduce overlapping assumptions or blind spots.
Clinical experts should establish the relevant risk categories and behavioral expectations. Synthetic scenarios can then be generated through complementary techniques, including structured simulation and steering. Finally, outputs should demonstrate reliability and clinical validity.
At Circuit Breaker Labs, our approach emphasizes clinically-informed risk criteria and reproducible evaluation methods rather than solely on one generative model to independently produce scenarios and another to judge the results. We believe this separation is an important design principle for developing trustworthy evaluation infrastructure.
The involvement of clinicians experienced with underrepresented populations is especially important. Available patient datasets may not fully reflect how individuals across age groups, dialects, health literacy levels, cultural contexts, and co-occurring conditions describe their concerns. Specialists and generalists can identify clinically meaningful variations that would otherwise be absent from test-generation processes.
We also recommend that synthetic datasets be continuously expanded and validated as new evidence emerges from clinical practice, postmarket monitoring, patient feedback, and independent evaluations.
The appropriate goal is therefore not to replace real-world evidence, but to extend the range of conditions under which safety can be examined before patients encounter those conditions themselves.
Question 14: Performance Comparators
What clinical standards, physician expertise, or human-AI performance benchmarks should determine whether an AI device performs acceptably?
We recommend that performance comparators and acceptance criteria be determined according to a device's intended use, level of autonomy, patient population, and clinical consequences of failure.
The proposed axis framework could provide a starting point for establishing baseline expectations. Manufacturers should define the functions within the device's intended scope, the behaviors that would constitute unacceptable deviations from that scope, and the clinical competencies required for safe operation.
Minimum requirements could include recognition of clinically significant contraindications, appropriate escalation in emergencies, adherence to the device's intended scope, and avoidance of clearly harmful recommendations. These should be grounded in applicable clinical standards and established with input from qualified specialists.
Comparative performance could then be assessed against clinicians or human-AI teams appropriate to the intended workflow.
The relative contributions of generalist and specialist physicians deserve particular consideration.
We therefore recommend involving both groups in evaluation design while assigning responsibility for defining particular clinical acceptance criteria to professionals with qualifications appropriate to the clinical question.
The degree of human involvement in analogous real-world operation should also determine which comparator is most relevant. If a system is intended to support a clinician who reviews and acts on its recommendations, the combined performance of the clinician and AI may be a central endpoint. If the system makes decisions or takes actions autonomously, its independent performance may be assessed.
Patient-facing systems present an additional consideration. The absence of professional review at the point of interaction may necessitate more conservative safety requirements, stronger escalation pathways, and ongoing monitoring. However, retrospective human review should not automatically be treated as equivalent to clinician supervision before a consequential response reaches a patient.
Finally, because generative systems can produce many different clinically acceptable responses, scoring should focus on whether the response satisfies clinically meaningful criteria rather than whether it matches a single reference answer.
Question 15: Real-World Alternatives
Should AI medical devices be evaluated against the care patients would realistically receive without them, including delayed care or no treatment?
We strongly support consideration of the care or intervention that would realistically occur in the absence of a generative AI-enabled device. This is particularly important for applications intended to improve access-to-care.
In behavioral health, for example, the relevant alternative to an AI-enabled support tool may not always be an appointment with a licensed therapist. For some individuals, the alternative may be delayed care, limited access to clinical information, reliance on informal support, or no intervention at all.
Evaluating a device exclusively against an idealized healthcare scenario may therefore underestimate its potential benefits, particularly for populations facing substantial barriers to access.
However, we recommend distinguishing benefit relative to the available alternative from the device's absolute safety. A device may offer meaningful benefits where no timely care would otherwise be available while still creating risks that require independent assessment.
One particularly important question is whether an AI interaction introduces a clinically significant harm that would not otherwise have occurred.
For example, in behavioral health, a user experiencing ordinary sadness may seek conversational support. An AI system that amplifies emotional distress, or encourages inappropriate dependence could potentially worsen the user's condition, even if it never provides an explicitly dangerous instruction.
Conversely, a well-designed system may provide helpful information, encourage appropriate care-seeking, or offer beneficial support during periods when professional care is unavailable.
We recommend that clinical confirmation consider whether the device produces a net benefit, neutral effect, or additional harm relative to a realistic counterfactual, while separately measuring serious adverse outcomes that should not be offset by average benefit.
Potential comparators could include usual care, delayed specialist review, unaided clinical judgment, available digital resources, or no intervention, depending on the intended use and actual deployment setting.
For devices performing clinical reasoning or decision-making, comparison with qualified clinicians remains important. Specialist judgment may be particularly relevant for assessing complex cases, while generalist judgment can help characterize the decisions likely to occur earlier in a patient's care pathway.
When feasible, studies could examine not only whether the device's decision agrees with clinical experts, but also whether exposure to the device changes patient understanding, decision-making, care-seeking behavior, or clinical outcomes.
Such assessments should account for baseline patient characteristics and other factors that may influence outcomes independently of the AI system.
We believe this approach would allow FDA to recognize the substantial potential of generative AI to expand access to healthcare without overlooking the possibility that some applications create new harms rather than simply performing existing clinical functions less effectively.
Question 16: Independent Evaluation
What role should qualified independent third parties play in AI medical device evaluation, and how can their participation preserve competition and innovation?
We strongly support the involvement of qualified independent third parties throughout the competency-based evaluation process. Blind spots are natural, and independent scrutiny is particularly valuable for identifying failure modes that may not be apparent to the organizations developing a device.
Independent evaluators can contribute during premarket development by identifying vulnerabilities, assessing clinically significant behaviors, and providing evidence that informs design decisions and safeguards. This need not be limited to a final assessment immediately before submission. Early and iterative evaluation can help manufacturers address risks while products are still being developed, rather than discovering them after deployment.
The resulting evidence can also establish a baseline for postmarket monitoring. By evaluating devices against consistent clinical factors over time, independent organizations can help manufacturers identify emerging vulnerabilities, quantify regressions, and determine whether modifications have affected previously demonstrated competencies.
We recommend that qualifications reflect both clinical and technical expertise. Evaluators should demonstrate the ability to define clinically meaningful failure categories, incorporate perspectives from relevant specialists and patient populations, conduct sufficiently broad and high-volume testing, and produce reproducible, auditable findings. Technical testing capacity alone should not be sufficient; evaluation methods must also demonstrate clinical validity.
Importantly, independent evaluation need not require access to a manufacturer's proprietary models, training data, system prompts, or internal architecture. For example, at Circuit Breaker Labs, we conduct external evaluations through API endpoint-based access to deployed systems, allowing manufacturers to retain control of their underlying intellectual property. This approach can reduce barriers to participation while enabling independent assessment of observable device behavior.
Finally, we suggest that CDRH avoid creating an evaluation framework that inadvertently concentrates market access among a small number of organizations. Qualified for-profit companies, nonprofit laboratories, academic groups, and other independent evaluators should be able to participate under transparent, technology-neutral qualification criteria.
Independence should be defined by the integrity of the evaluation process, not the legal or commercial structure of the organization conducting it. Such an approach would encourage competition in evaluation methodologies, encourage innovation in the evaluation process itself, protect manufacturer intellectual property, and support the development of safety infrastructure capable of keeping pace with innovation.
Question 17: Multimodal AI
How should competency-based evaluation adapt to medical AI systems that process voice, images, and other modalities beyond text?
We believe the competency-based approach can be applied across different model architectures, provided that the competencies are defined according to the device's intended clinical function rather than the particular architecture of its underlying model.
The central regulatory question should remain whether the final deployed device performs its intended function safely and effectively under foreseeable conditions. However, multimodal devices introduce additional failure modes that may not be adequately captured through evaluations designed primarily for text-based systems.
For example, a voice-enabled device may respond appropriately to an accurate transcript while failing to recognize clinically significant information when it is spoken with an unfamiliar accent, interrupted, or affected by background noise. We have observed this with our voice red-teaming offering at Circuit Breaker Labs. Similarly, an image-based system may accurately interpret a clear clinical image but behave differently when the image is incomplete, poorly illuminated, or inconsistent with accompanying patient information.
We therefore recommend evaluating both individual components and the complete end-to-end clinical interaction. Component-level testing can help identify whether a failure originated in speech recognition, image interpretation, retrieval, reasoning, or another process. However, successful performance by individual components does not necessarily establish that the complete device will behave safely when those components interact.
An especially important competency is the handling of conflicting information across modalities. For example, a patient's verbal description of a condition may conflict with an image, a physiological measurement, or a previously documented clinical history. Evaluation should establish whether the device appropriately recognizes uncertainty, gathers additional information, or defers when necessary rather than arbitrarily prioritizing one input source.
Multimodal evaluation should also consider patient-specific communication patterns, including accents, dialects, age-related language differences, interrupted speech, and changes in how users present information across the course of an interaction.
Multi-participant red-teaming, an offering Circuit Breaker Labs, is also critical to consider as some AI-enabled medical devices may involve one or more patients and clinicians interfacing with an AI product, where risk is between human-AI and human-human.
For generative models, additional considerations may include the accuracy of modeled state transitions, the system's response to uncertainty, and the consequences of incorrect predictions over multiple steps. A plausible simulated future state does not necessarily establish that the predicted outcome is clinically reliable.
These systems should be assessed through realistic scenarios in which initial information may be incomplete, assumptions may be incorrect, and errors can accumulate over time.
Postmarket Monitoring
Question 18: Premarket vs. Postmarket Evidence
Under what circumstances should the FDA accept greater uncertainty before approval in exchange for stronger postmarket monitoring?
We recognize the rationale for considering greater reliance on postmarket monitoring, particularly because generative AI-enabled medical devices may operate in settings where real-world usage patterns and the full distribution of clinical risks are not yet well characterized.
However, postmarket monitoring should complement a meaningful premarket evidence baseline. This is particularly important for emerging patient-facing applications. In behavioral health, for example, available real-world data may not adequately represent the range of ways patients discuss sensitive conditions. Privacy constraints, limited data availability, and the evolving nature of human-AI interactions can make it difficult to construct comprehensive evaluation datasets from historical conversations alone.
We therefore recommend that manufacturers use clinically-informed synthetic scenarios to supplement available real-world evidence during premarket evaluation. These scenarios can systematically explore clinically plausible variations in patient characteristics, language, symptom presentation, comorbidities, and conversational trajectories.
High-volume testing is especially valuable in this context because it enables manufacturers to examine a broad range of foreseeable failures before users encounter them. Clinical experts should be involved in defining the risks being tested and evaluating whether synthetic scenarios reflect meaningful clinical situations.
Postmarket monitoring can then assess whether performance remains consistent with that baseline, whether changes introduce new vulnerabilities, and whether previously unidentified failure modes emerge as real-world evidence accumulates.
Greater reliance on postmarket evidence may be reasonable when the device has demonstrated adequate initial safety, the remaining uncertainties can be monitored meaningfully, and mechanisms exist to identify and remediate clinically significant problems in a timely manner.
For patient-facing conversational systems, human psychology introduces further considerations. A device's outputs may influence trust, emotional reliance, care-seeking behavior, and subsequent disclosures. These effects may develop across extended interactions and should not be assumed to be captured by conventional single-turn benchmarks.
We therefore recommend an approach in which robust premarket synthetic and clinical evaluation establishes a safety baseline, while continuous monitoring expands the evidence base and detects deviations after deployment.
This would support innovation without treating patients as the primary means of discovering risks that could reasonably have been identified before market entry. Humans should not be Guinea pigs.
Question 19: Postmarket Monitoring
How frequently should medical AI devices be reassessed, and what changes or events should trigger additional safety testing?
We support periodic re-benchmarking, sample-based clinician review, and performance degradation monitoring. However, we recommend that CDRH explicitly distinguish event-triggered reassessment from time-based reassessment, and consider both essential components of postmarket oversight.
Event-triggered evaluations are valuable when manufacturers make known changes that may affect device performance. Examples include modifications to the underlying foundation model, prompts, retrieval architecture, safety controls, guardrails, or clinical functionality.
However, manufacturers may not always know when their systems have changed in clinically significant ways. Third-party model providers may modify hosted services, previously unrecognized vulnerabilities may emerge, and changes in deployment conditions may affect behavior without a clearly identifiable software modification.
For this reason, postmarket monitoring should not depend exclusively on manufacturers recognizing that a change has occurred.
We recommend recurring re-benchmarking against established clinical safety criteria, even when no known modification has taken place. Such evaluations could identify clinically meaningful regressions, newly emerging vulnerabilities, or changes in the consistency of safety-critical behavior.
The appropriate cadence should be proportionate to the device's risk profile, anticipated usage, clinical setting, and rate of technical change. Monthly or quarterly evaluation may be appropriate for some applications, while higher-risk functions may require more frequent monitoring.
Clinical expertise should remain central throughout this process. Clinicians should contribute to defining evaluation criteria, identifying emerging clinical risks, and adjudicating cases where automated methods cannot reliably determine the clinical significance of an output.
We further recommend that monitoring include refreshed or independently developed challenge cases to help detect vulnerabilities that may not appear in previously used benchmarks.
The objective should be to create a monitoring system that identifies both regressions in known competencies and failures that were not anticipated during initial evaluation.
Question 20: Automated AI Oversight
How should the FDA evaluate the accuracy, reliability, and limitations of automated systems used to monitor other medical AI devices?
Machine-based supervisory systems can absolutely play an important role in scalable postmarket monitoring, particularly where devices produce large volumes of open-ended outputs and tests are necessary for generative AI systems, but cannot feasibly be reviewed individually by clinicians.
However, we caution against assuming that one generative model can reliably validate another. LLM-as-a-judge approaches exhibit several documented limitations, including:
These concerns are especially significant when supervisory systems are used to determine whether a safety-critical behavior has improved, degraded, or remained unchanged.
A broader concern is correlated error. When the device under evaluation and the supervisory model share similar assumptions, training influences, or failure modes, agreement between the two may create a false impression of independent validation.
We therefore recommend that CDRH encourage the use of complementary and independently validated evaluation mechanisms, including non-generative methods, specialized scoring systems, and qualified clinician review where appropriate.
Non-generative evaluation techniques can be designed to produce reproducible numerical scores while assessing clinically relevant contextual features. Circuit Breaker Labs, for example, develops clinically-informed machine-based evaluation methods that do not rely on a general-purpose LLM to assign final judgments.
We recommend that manufacturers characterize the reliability of any supervisory evaluation system by requesting repeated assessments of identical outputs.
CDRH should also consider evaluation throughput and operational scalability. A central advantage of machine-based supervision is the ability to evaluate substantially larger volumes of interactions than manual review alone. However, this advantage is meaningful only when the evaluation method is sufficiently accurate, reliable, and clinically appropriate.
Ultimately, the supervisory system should itself be treated as a measurement instrument whose validity, reliability, and limitations must be demonstrated. A changing device cannot be meaningfully monitored using an evaluation method whose own behavior is insufficiently characterized.
Question 21: Stakeholder Accountability
How should clinicians, healthcare institutions, independent evaluators, and other stakeholders contribute to postmarket monitoring while keeping manufacturers accountable?
All these stakeholders have valuable roles in postmarket monitoring, but CDRH should also recognize the important contributions of independent, for-profit evaluation organizations. Given the pace at which generative AI is advancing, a sustainable safety ecosystem must be capable of innovating at a comparable pace to for-profit manufacturers. Commercial evaluation organizations can have incentives to invest in scalable testing infrastructure, develop new methodologies, and rapidly adapt to emerging risks and new capabilities. Their involvement should be encouraged alongside academic institutions, professional societies, clinicians, and other stakeholders.
Shared participation in monitoring need not mean shared accountability for remediation. Manufacturers can remain responsible for defining their device's risk profile and demonstrating that appropriate safeguards are in place. Independent evaluators and other stakeholders can supplement this work by identifying previously unrecognized vulnerabilities, assessing performance against established clinical criteria, and detecting emerging risks through postmarket monitoring and reporting those back to manufacturers.
When monitoring identifies a clinically significant issue, the manufacturer should be responsible for investigating and addressing it within a timeframe proportionate to its severity. Where remediation cannot be completed within that timeframe, manufacturers should document a corrective action plan, interim risk mitigations, and an expected resolution timeline, with reporting to FDA where applicable. Particularly serious risks may require immediate intervention rather than awaiting completion of a remediation plan.
This approach allows different stakeholders to contribute their expertise without obscuring responsibility. Independent organizations can identify and characterize failures, clinicians and professional societies can inform their clinical significance, and manufacturers remain accountable for maintaining the safety and effectiveness of their devices.
Question 22: Re-Benchmarking After Changes
How should the extent of safety testing and regulatory review depend on the type and potential consequences of modifications to an AI device?
We recommend that post-deployment reassessment be proportional to the potential clinical consequences of a modification rather than solely its technical complexity or the manufacturer's intended purpose for the change.
A relatively minor software modification may have outsized implications for device safety. For example, changes to a system prompt, knowledge retrieval mechanism, context management strategy, or underlying foundation model may alter how a device interprets contraindications, recognizes clinical emergencies, or responds to patients outside its intended scope.
We propose a risk-based re-benchmarking framework built on the premarket competency assessment.
Manufacturers should maintain a mapping between their device's components, intended functions, clinically-significant failure categories, and established acceptance criteria. When a modification occurs, this mapping can help identify the competencies potentially affected and the evidence needed to demonstrate continued performance.
For example, a change to the user interface that does not materially affect clinical information or user decision-making may warrant limited reassessment. A change to the underlying language model, safety architecture, or conversational memory system may warrant broader testing, including multi-turn scenarios and relevant clinical subpopulations.
However, even targeted reassessment should be supplemented by cadenced safety-critical regression tests. Generative systems can exhibit unexpected interactions between components, and manufacturers should not assume that a modification affects only the functions they intended to change.
We recommend distinguishing regression testing from discovery-oriented testing. Regression testing determines whether established competencies have been maintained. Discovery-oriented testing uses refreshed, independently developed, or adversarial scenarios to identify new vulnerabilities that may emerge unexpectedly.
Changes that fall within a prospectively defined and appropriately validated modification framework may be suitable for management through a PCCP, consistent with applicable regulatory requirements. Changes that introduce new intended functions, substantially alter risk exposure, or exceed the boundaries of an established PCCP may require additional evidence or FDA review.
Importantly, the amount of evidence should be informed by the severity and detectability of potential harm, the degree of clinical autonomy, and the availability of effective downstream safeguards.
This approach would make re-benchmarking more efficient and predictable while recognizing that the safety impact of a modification cannot always be inferred from the size or apparent significance of the underlying engineering change.
Question 23: Unpredictable Modifications
How should predetermined change control plans (PCCPs) account for AI system changes that cannot be fully anticipated in advance?
PCCPs are valuable for modifications manufacturers can reasonably anticipate, including changes to underlying models, prompts, retrieval systems, or other components that may affect device performance. However, generative AI-enabled devices present an additional challenge: clinically meaningful behavioral changes may occur even when the manufacturer has not knowingly implemented a modification.
For example, a third-party foundation model provider may update a hosted model without providing sufficient visibility into the change. Even when the model identifier remains unchanged, differences in backend configuration or other dependencies may potentially affect performance.
We therefore recommend complementing modification-triggered PCCPs with proactive, recurring evaluation. Rather than testing only when a manufacturer anticipates that something has changed, a device could be periodically reassessed against its established clinical competencies and safety requirements, regardless of whether a modification has been identified.
This would create two complementary safeguards: predefined change-control procedures for anticipated modifications, and ongoing monitoring to detect unexpected changes, previously unidentified vulnerabilities, or performance degradation.
Importantly, the monitoring framework should define which safety-critical behaviors must remain consistent, how changes will be measured, what constitutes a clinically significant regression, and what corrective actions would follow. Evaluation methods should produce numerical, sufficiently reproducible results that can be tracked to distinguish changes in device performance from variability in the measurement process itself.
This approach would allow manufacturers to prepare for the reality that not every future change can be predicted, without requiring them to prespecify every possible modification. The central principle is that manufacturers may not know what will change, but they can define what safety-critical behaviors must continue to hold.
Question 24: Third-Party Model Updates
How can medical device manufacturers detect and address safety risks introduced by updates to third-party foundation models?
Third-party foundation model dependencies create a significant challenge because manufacturers may not control, anticipate, or even be notified of changes that affect their device's clinical performance.
We recommend that manufacturers implement ongoing behavioral monitoring and comparative evaluations of their deployed systems. Independent evaluation organizations, including companies such as Circuit Breaker Labs, can support this process through recurring safety assessments and controlled comparisons, or bakeoffs, between existing and prospective model configurations. For example, when a manufacturer considers replacing an underlying foundation model, the current and proposed versions of the complete device can be evaluated against common clinical safety criteria, including relevant failure modes and population-specific requirements. Because Circuit Breaker Labs produces numerical, explainable results, manufacturers can identify potential improvements and regressions before implementing the change.
The same approach can help detect unexpected changes introduced by third-party model providers. By periodically evaluating the deployed device against an established baseline, manufacturers can identify behavioral regressions even when they have not received notification of an underlying modification.
We believe recurring evaluation should be a reasonable expectation for GenAI-enabled medical devices, with frequency proportionate to clinical risk and the pace of system change. Monthly or quarterly assessments may be appropriate for some applications, while high-consequence systems or those dependent on frequently updated foundation models may warrant more frequent or event-triggered testing.
Importantly, periodic testing should complement, rather than replace, evaluations conducted before known model changes are deployed. Manufacturers should also consider technical and contractual mechanisms such as model-version pinning, advance update notifications, access to previous versions, and rollback capabilities where available.
Advances in automated, clinically-informed evaluation make broad regression testing substantially more scalable than exclusively manual approaches. This creates an opportunity to maintain rigorous oversight without imposing the same cost and operational burden as repeated full-scale clinical studies.
Ultimately, the appropriate regulatory expectation should not be that manufacturers can prevent every third-party model change. It should be that they maintain reliable mechanisms to detect clinically significant changes, assess their consequences, and take timely corrective action, regardless of whether the change originated within or outside their organization.
Other Topics
Question 25: Foundation Model Documentation
Are voluntary Foundation Model Master Files (MAFs) sufficient to support regulatory review, or should the FDA consider alternative evidence mechanisms?
We appreciate the potential value of voluntary Foundation MAFs, particularly as a mechanism for improving transparency about underlying models used in medical devices. However, we are concerned that voluntary disclosure alone may not provide sufficiently current or application-specific evidence of safety.
Foundation model developers may be slow to continually update regulatory documentation, particularly when they are not themselves developing the downstream medical device and may not be attuned to the specific risks.
A more fundamental limitation is that foundation model safety is not equivalent to device safety. A foundation model's performance depends substantially on how it is incorporated into a specific application. System prompts, retrieval-augmented generation, conversational memory, and guardrails can all introduce new vulnerabilities or alter existing ones.
Similarly, the clinical safety requirements for a behavioral health application may differ substantially from those for a medication management system, even when both use the same underlying foundation model.
For these reasons, we would not recommend relying on a voluntary MAF as the primary mechanism for establishing or maintaining the safety of downstream devices.
Instead, CDRH could consider a complementary framework involving ongoing independent foundation model evaluations and application-specific testing.
For example, qualified independent evaluation organizations could periodically compare major foundation models against common healthcare-relevant competencies, including clinical knowledge, safety-critical recognition, uncertainty communication, and consistency of behavior.
Such comparative evaluations, “bakeoffs” as Circuit Breaker Labs refers to them, could provide useful evidence about the relative strengths, limitations, and changing performance of different underlying models.
CDRH could consider maintaining or referencing a body of independently-generated evaluation evidence, with clear documentation of model versions, assessment methods, intended contexts, and known limitations. This could provide reviewers with additional information when assessing devices built on commonly used foundation models.
At Circuit Breaker Labs, comparative evaluation of different model configurations is one approach we use to help characterize clinically relevant performance differences. We believe a similar methodology could contribute to a broader evidence ecosystem without requiring foundation model developers to disclose proprietary training information or pending non-public updates.
However, these comparisons should not be interpreted as general authorization or certification of an underlying model for medical use. A model that performs well on common healthcare benchmarks may still behave unsafely when integrated into a particular device or altered via RAG, etc.
If this approach is taken, we recommend treating foundation model evidence as a useful reference that can inform, but not replace, application-specific evaluation.
Voluntary MAFs could still contribute important information, including version identification, known limitations, major safety-relevant changes, and update-notification practices. These disclosures would be most useful when combined with proactive, recurring behavioral testing rather than relied upon as a static representation of model safety.
The objective should be to establish a regulatory evidence framework that remains informative even when model providers make unanticipated changes, use cases evolve, or the performance characteristics of a previously evaluated model shift over time.
Question 26: Agentic AI
What additional safety requirements and oversight are necessary for autonomous medical AI systems that perform multi-step actions and use external tools?
Agentic generative AI-enabled devices warrant additional scrutiny because their risks may emerge across sequences of decisions and actions rather than from any individual output.
This is particularly important for systems capable of using external tools, maintaining conversational context, modifying records, initiating clinical workflows, or performing actions with limited opportunities for human review.
We recommend that CDRH place particular emphasis on high-volume, clinically-informed, trajectory-level evaluation in both premarket and postmarket settings.
Generative AI systems can produce variable outputs in response to similar inputs, and agentic systems introduce further complexity because small differences in an early decision may propagate through subsequent actions. Evaluating a small number of isolated examples is therefore unlikely to adequately characterize the range of potential failures.
Manufacturers should evaluate realistic multi-step scenarios, including variations in the order of information, incomplete patient histories, conflicting inputs, changing clinical conditions, and extended interactions in which important safety information may become less salient.
A central concern is context degradation. An agent may initially recognize a clinically significant contraindication but subsequently fail to account for it when performing an action many steps later. Likewise, an agent may correctly execute individual tool calls while producing an unsafe overall outcome because it has lost track of the clinical context or failed to reassess whether its planned actions remain appropriate.
These risks are not hypothetical properties of tool use alone. They arise from the interaction between the underlying generative system, its context-management mechanisms, the tools it can access, and the clinical consequences of its decisions.
Our experience conducting trajectory-based evaluations in behavioral health has reinforced the importance of assessing complete interactions. The order in which information is introduced, the progression of the user's circumstances, and the accumulated conversational context can materially affect how safety safeguards are invoked and whether a system maintains appropriate boundaries.
We recommend that acceptance criteria therefore address not only the correctness of individual responses or tool calls, but also the safety of the complete action sequence.
Relevant competencies should include persistent recognition of clinically significant information, adherence to authorized scope, appropriate human-approval checkpoints, resistance to prompt injection, handling of erroneous tool outputs, and the ability to stop or safely recover when circumstances change.
Equally important is the scientific reliability of the evaluation process itself. Agentic systems produce complex, open-ended trajectories, but this does not mean their performance must be assessed through subjective or irreproducible judgments.
Evaluation methods can incorporate clinical context and nuanced behavioral criteria while producing structured, quantitative, and auditable results. We recommend that manufacturers demonstrate how their scoring methods distinguish clinically significant failures, how reliably those methods agree with qualified clinical judgment, and whether repeated evaluation of identical trajectories produces consistent results.
A zero-variance evaluator should assign the same score to an identical trajectory. Such reproducibility is particularly valuable for identifying safety regressions across device versions and postmarket updates. Generative LLM-as-a-judge approaches may provide useful qualitative assessments, but their variability, systematic biases, and potential correlated errors should be characterized before they are relied upon for safety-critical determinations. A changing evaluation method can make it difficult to determine whether a device has actually improved or deteriorated.
Finally, we recommend that agentic device evaluation combine large-scale synthetic testing with clinically representative confirmation and continuing postmarket reassessment. Testing should be sufficiently broad to explore complex and uncommon trajectories, while statistical claims should account for the representativeness and dependence of generated scenarios.
The objective should be to establish reproducible evidence that clinically significant safeguards persist across complex action sequences, changing contexts, and evolving system configurations.
As generative AI becomes increasingly capable of acting rather than merely providing information, the importance of measuring these properties will only grow.

Work with Us
Prove Your AI
Does No Harm.